Tests and Adjustments
Adjustments
Equipment Used for the Source Adjustment
Equipment Type
Power meter
Power sensor, Type-N
Power sensor, 2.4 mm
Power sensor, V-Band
Adapter, WR-15 to 1.85 mm (m)
Adapter, 1.85 mm (f) to 1.85 mm (f)
Adapter, 2.4 mm (f) to 2.4 mm (f)
Adapter, Type-N (f) to 2.4 mm (f)
a. This sensor and adapter must be calibrated together by the Keysight factory. The data supplied is
b. The 1.85 mm connector has the same ruggedness and is compatible with the 2.4 mm connectors.
Procedure
1. Connect the equipment as shown in
between the network analyzer and the power meter.
Figure 3-17
Equipment Setup for the Source Adjustment
2. .Press UTILITY
click Source Adjustment.
3. Ensure the GPIB settings are correct.
Keysight N5247B Service Guide
a
b
only valid as long as the sensor and adapter remain connected.
1.85 mm and 2.4 mm adapters are interchangeable.
, then
System
Model or
Part Number
E4418B/E4419B
8482A
8487A
V8486A
V281B
85058-60114
11900B
11903B
Figure
3-17. Connect a GPIB cable
Service
Adjustment Routines
, then
Alternate Model or
Part Number
E4418A/E4419A
None
None
None
None
None
85056-60007
None
, then