Keysight Technologies N5247B Service Manual page 56

2-port and 4-port pna-x microwave network analyzer
Hide thumbs Also See for N5247B:
Table of Contents

Advertisement

Tests and Adjustments
System Verification
If the system verification is performed with a non-Keysight cable, ensure that
the cable meets or exceeds the specifications for the test cable specified in the
previous table,
to the cable's user's guide for specifications.
Kit Substitution
Non-Keysight calibration kits and verification kits are not recommended nor
supported.
System Verification Procedure
1. If you desire printed test outputs, connect a printer to the analyzer. For the
printer, ensure that the correct driver is loaded and the printer is defined
as the default printer. Refer to the embedded help in the analyzer for
printer setup. Let the analyzer warm up for at least 90 minutes.
2. Insert the verification kit disk into the analyzer disk drive.
3. Press UTILITY
System Verification dialog box is displayed; refer to
Figure 3-10
System Verification Dialog Box
4. In the Calibration Kit box, select the calibration kit or electronic calibration
module (ECal) that is being used by clicking on it. The corresponding
verification kit to use is selected for you and displayed in the Verification
Kit box. Refer to
5. Under Printer Output, click one of the following options. Refer to
3-10.
3-26
"Equipment Used in the System Verification Procedure."
Service
System
, then
Figure
3-10.
System Verification
, then
Figure
3-10.
Keysight N5247B Service Guide
Refer
. The
Figure

Advertisement

Table of Contents
loading

Table of Contents