Keysight Technologies N5247B Service Manual page 27

2-port and 4-port pna-x microwave network analyzer
Hide thumbs Also See for N5247B:
Table of Contents

Advertisement

General Product Information
Required Service Test Equipment
a
Equipment
Calibration and Verification Tools
1.85 mm calibration kit
1.85 mm verification kit
ECal module
Cables
BNC cable (2 required)
1.85 mm RF cable (Qty 2)
GPIB cable
Adapters
1.85 mm (f) to 1.85 mm (f)
d
2.4 mm (f) to Type-N (f)
Attenuators
1.85 mm (m, f), 10-dB fixed
attenuator
1.85 mm (m, f), 20-dB fixed
attenuator
a. Unless specified otherwise, equipment listed is required for all analyzer models.
b. P = Performance tests, A = Adjustments, T = Troubleshooting, R = Repair, V = System verification
c. Included in the 85058B/E calibration kits.
d. The 1.85 mm connector has the same ruggedness and is compatible with the 2.4 mm connectors. 1.85 mm and 2.4 mm
adapters may be used interchangeably.
a
Equipment
Tools
T-8 TORX driver
T-10 TORX driver
Keysight N5247B Service Guide
Critical Specifications
--
--
--
50 Ω, length ≥ 60 cm
50 Ω, length ≥ 60 cm
N/A
Return Loss: ≥ 20 dB
Return Loss: ≥ 28 dB
Freq: DC to 67 GHz
Freq: DC to 67 GHz
Critical Specifications
0.6 N-m (5 in-lb) setting
0.5, 0.8, and 1.0 N-m
(4, 7, and 9 in-lb) settings
Recommended
Alternate Model
Model or Part
or Part Number
Number
85058B
85058E
DC to 67 GHz
DC to 67GHz
85058V
None
45 MHz to 67 GHz
N4694A
None
10 MHz to 67 GHz
8120-1839
None
N4697E
Equivalent
10833A/B/C/D
None
None
c
N5520B
11903B
None
8490G
None
Option 010
8490G
None
Option 020
Recommended
Alternate
Model or Part
Model Number
Number
N/A
N/A
N/A
N/A
b
Use
P,T
V
V
A
P,A,V
P,A
P,A,T
P,A,T
P
P
b
Use
R
T,R
2-5