5
Input Pressure value, and press [ENT].
[Example] Pres. :1020.0hPa.
The display returns previous mode.
*1) *2)
*1) Range : Temp.
-30.0 ° C ~ +60.0 ° C (0.1 ° C step)
Pres.
315.0 ~1066.0hPa (0.1hPa step) , 420.0 ~800.0mmHg (0.1mmHg step)
16.5 ~ 31.5inHg (0.1inHg step)
*2) When the atmospheric correction value, which is calculated from the input temperature and pressure
values, exceeds the range ± 999.9ppm, the operating procedure returns to step
values again.
How to Set the Atmospheric Correction Value Directly
Measure the temperature and air pressure to find atmospheric correction value(PPM) from the chart or
correction formula.
Operating procedure
1
Press STAR ( ) key.
2
Press [F6] ( ↓ )key to get the function as in
page 2.
3
Press [F3] key.
Current setting value is displayed.
4
Move the cursor (>) to PPM setting by
pressing [F5](→) key.
5
Enter atmospheric correction value and press
[ENT] key. *1)
The display returns previous mode.
*1) Input range : -999.9mm ~ +999.9mm, 0.1mm step
10 SETTING ATMOSPHERIC CORRECTION
Enter
Pres.
[ENT]
1012.0hPa
Pressure setting
Operation
[ ]
[F6]
[F3]
1012.0hPa
Move
cursor
Enter
PPM
[ENT]
1012.0hPa
10-2
5.1ppm
20.0 °C
0.0 mm
3
automatically. Input
Display
2001-07-10 14:30:40
3
1
2001-07-10 14:30:40
2
5.1ppm
20.0 °C
0.0 mm
PPM setting
5.1ppm
20.0 °C
+14.0 mm