Care instructions
ESTETICA E50 Life

Advertisement

Table of Contents
loading

Summary of Contents for KaVo ESTETICA E50 Life

  • Page 1 Care instructions ESTETICA E50 Life...
  • Page 2 Manufacturer: Distributed by: Kaltenbach & Voigt GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach www.kavo.com Phone +49 7351 56-0 Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Target group ...................... 4 2 Setup methods according to ISO 17664 ................5 2.1 References........................5 2.2 General instructions for the reprocessing of KaVo treatment units........6 2.2.1 Overview ESTETICA E50 Life reprocessing............. 7 2.3 Reprocessing of the surfaces of treatment unit / upholstery ..........8 2.4 Preparation of the dentist unit ..................
  • Page 4: User Instructions

    1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. Note Comply with the Instructions for Use of the ESTETICA E50 Life treatment unit! 1.1.1 Abbreviations Explanation brevi- ation...
  • Page 5: Setup Methods According To Iso 17664

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.1 References 2 Setup methods according to ISO 17664 2.1 References The reprocessing consists of the following basic steps: ▪ Cleaning and disinfection ( manually or automatically) ▪...
  • Page 6: General Instructions For The Reprocessing Of Kavo Treatment Units

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units 2.2 General instructions for the reprocessing of KaVo treatment units Please comply with the general reprocessing instructions at all times. Detailed product-specific reprocessing instructions are provided subsequently.
  • Page 7: Overview Estetica E50 Life Reprocessing

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units Manual drying ▶ Supply dry, clean and uncontaminated compressed air according to ISO 7494-2 only. ▶ Blow off the outside and inside with compressed air until water drops are no longer visible.
  • Page 8: Reprocessing Of The Surfaces Of Treatment Unit / Upholstery

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit / upholstery Reprocessing items/medical Cleaning Cleaning Disinfec- Disinfec- Sterilisa- devices manual automatic tion tion tion manual automatic automatic Assistant element...
  • Page 9: Preparation Of The Dentist Unit

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Approved disinfectants: ▪ CaviCide / CaviWipes (Metrex) ▪ Microcide AF Liquid (Schülke & Mayr) ▪ FD 322 (Dürr Dental) ▪ Incidin Liquid (Ecolab) ▶...
  • Page 10 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Cleaning and disinfection of S-table handle and holder pad/anti-slip mat ▶ Press the push button and pull off the handle. ▶ Clean the handle under flowing water.
  • Page 11: Reprocessing Of The Instrument Holder Of The Tm Table

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit 2.4.2 Reprocessing of the instrument holder of the TM table Note The handpiece holder (Mat. no. 1.004.4838) can be removed for better cleaning.
  • Page 12: Reprocessing Of The Instrument Arm S Table

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Pull the snap-in buttons outward evenly and carefully remove the instru- ment holder. ▶ Clean the instrument holder under running water.
  • Page 13 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Attach the germ reduction attachment. with patient part assembly kit / without patient part assembly kit ▶ Take the handpieces off the hoses.
  • Page 14: Setting Up The 3-Function And Multifunction Handpiece

    Note Wet the O-rings of the gripping sleeves after sterilisation using silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replacing the cannula, the air channel must be blown out before the start of treatment to ensure that the air exiting during treatment is dry.
  • Page 15 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Clean the key part, gripping sleeve and cannula by brushing them under running water (at least tap water quality). ▶ Immediately remove contamination from materials used in the dental prac- tice (impression materials, caustic chemicals).
  • Page 16 Requirement The multifunctional handpiece has been reprocessed. ▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510) to the O-rings of the joint between the gripping sleeve and cannula. Use cotton buds for this purpose.
  • Page 17: Preparation Of The Instrument Hoses

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit NOTICE Damage caused by over-heating. Damage to the gripping sleeve. ▶ Immediately remove the parts from the steriliser after the sterilisation cycle is completed.
  • Page 18: Reprocessing Of The Kit For Physiological Saline Solution

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Note The turbine return air filter is in the same position in the dentist element vari- ant with S table. 2.4.7 Reprocessing of the kit for physiological saline solution The kit for physiological saline solution is optional.
  • Page 19 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Remove the old hose set sterile and dispose of it. ▶ Make sure that the flexing system is horizontal (turn manually if necessary).
  • Page 20: Reprocessing Kavo Ergocam One

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Open the hose clamp ⑦ before startup. Sterilisation kit for physiological saline solution Note The hose set sterile (10 pcs.) (Mat. no. 1.012.4601) is designed for single use and must be changed after each application.
  • Page 21 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Cleaning Note Disinfect the camera before cleaning. Disinfect after cleaning if necessary. Exterior cleaning by hand NOTICE Damage to the camera due to improper cleaning.
  • Page 22: Reprocessing Kavo Piezoled

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Approved disinfectants: ▪ MIKROZID Pump Spray - manufacturer: Schülke & Mayr ▪ Dürr System Hygiene FD 322 Spray disinfection - manufacturer: DÜRR Dental GmbH ▪...
  • Page 23 ▶ For programme settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the washer disinfector. ▶ In order to prevent negative effects on the KaVo medical device, make sure that the interior and the exterior of the medical device are dry after comple- tion of the cycle, and then lubricate it immediately with care products from the KaVo care system.
  • Page 24 For times and concentrations, please refer to the instructions of the manufac- turers of the cleaning/disinfection agent. Manual external disinfection KaVo recommends the following products based on compatibility of the materi- als. The microbiological efficacy must be confirmed by the disinfectant manu- facturer.
  • Page 25 ▶ For programme settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the washer disinfector. ▶ In order to prevent negative effects on the KaVo medical device, make sure that the interior and the exterior of the medical device are dry after comple- tion of the cycle, and then lubricate it immediately with care products from the KaVo care system.
  • Page 26 Use of worn tips Ultrasound tips are subject to wear and tear and become shorter during use. Worn tips are less effective and might cause pain to the patient. KaVo recommends using the components only for the specified service life. 26 / 58...
  • Page 27 Sterilisation Sterilisation of handpieces NOTICE Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products. NOTICE Contact corrosion due to moisture. Damage to product. ▶ Immediately remove the product from the steam steriliser after the steril-...
  • Page 28 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Place the cleaned and disinfected handpieces separately in sterilisation packages and weld them to be sealed, (e.g. KaVo STERIclave bags Mat. no. 0.411.9912). 135°C Steriliser with a triple pre-vacuum for at least 4 minutes at 134 °C ±...
  • Page 29: Reprocessing Of The Assistant Unit

    ▶ Place the cleaned and disinfected tips, endo files, file holders, endo wrenches and torque wrenches separately in sterilisation packages (e.g. KaVo STERIclave bags Mat. no. 0.411.9912) and weld them to be sealed or sterilise them in a sterilisation cassette (e.g. KaVo sterilisation cassette Mat.
  • Page 30: Reprocessing Of Sieve Inserts And Suction Hoses

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit Cleaning and disinfection of the suction hose connector ▶ Clean the disassembled parts under running water. ▶ Thermodisinfect the disassembled parts for at least 10 minutes at 93 °C...
  • Page 31: Preparation Of The Tumbler Holder And Filler

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit 2.6.2 Preparation of the tumbler holder and filler NOTICE Water in the device. Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl flush"...
  • Page 32: Preparation Of The Spittoon

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit The tumbler filler can be subjected to thermodisinfection. Sterilisation of the tumbler holder and filler NOTICE Damage caused by sterilisation. ▶ Do not sterilise the tumbler holder and filler.
  • Page 33: Manual Reprocessing Of The Water-Conducting System

    (Mat. no. 0.489.3451). ▶ If the Water bottle kit is used with the enclosed dosing attachment (Mat. no. 1.002.0287), add the proper amount of KaVo OXYGENAL 6 (Mat. no. 0.489.3451) with each filling. For the correct amount, please refer to the instructions of the dosing attachment for water germ reduction.
  • Page 34: Rinsing Programme

    During normal operation of the treatment centre, the continuous germ reduc- tion function secures continuous germ reduction in the water conducting sys- tems due to the automatic dosing of the germ reduction liquid, KaVo OXY- GENAL 6. Rinsing programme and intensive germ reduction facilitate the rinsing and germ reduction of the water-supplying system.
  • Page 35 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system Place the instruments on the germ reduction attachment with patient part assembly kit / without patient part assembly kit The cannula of the triple-function and multifunctional handpieces must be re- moved for the rinsing program and the intensive germ reduction.
  • Page 36 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system ▶ Insert the triple-function or multifunctional handpieces without the cannula into the clamping device of the germ reduction attachment. Make sure that the "W"...
  • Page 37: Intensive Germ Reduction

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system The rinsing program proceeds in four steps. For rinsing programme 01, the cycle starts in step 2, and for rinsing programme 02, the cycle starts in step 4.
  • Page 38 During intensive germ reduc- tion, the afferent water ducts are automatically filled with a higher concentra- tion of KaVo OXYGENAL 6. The intensive germ reduction program lasts ca. 45 minutes. The increased hydrogen peroxide concentration acts in the system for at least 30 minutes.
  • Page 39 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system Step Description Germ reduction of the triple-function and multifunctional handpiece and the spittoon bowl Duration: 60 seconds Germ reduction of the handpieces...
  • Page 40 The unit is now ready for use. Additional intensive germ reduction during period of non-use KaVo recommends carrying out an additional intensive germ reduction at uni- versities during semester breaks. This flushes the parts of the unit and replaces the germ reduction liquid.
  • Page 41: Reprocessing And Automatic Hygiene Functions Of The Suction System

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- 2.9 Reprocessing and automatic hygiene functions of the suction system 2.9.1 Reprocessing of the suction system including...
  • Page 42: Manual Internal Cleaning Of The Suction Hoses At The Extraction Point (Optional)

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- ter (limited rinsing). The water is aspirated from the Aquamat through the filter housing to the separation system. The spittoon bowl rinsing is alternately turned on and off.
  • Page 43 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- ð An automatically dosed amount of water is supplied by the vacuum. An acoustical signal is emitted after the process is completed (approx. 3 seconds).
  • Page 44 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- Preparation ▶ Fit the hose cleaning attachment (Mat. no. 0.763.8021) to the tap of the Aquamat. E50 Life including CENTRAmat / DEKAmat assembly kit ▶...
  • Page 45: Preparation Of The Amalgam Separator

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.10 Preparation of the amalgam separator Starting the tube washing function ▶ Press the "HYDROclean" button. ð The tube washing function starts. Note DEKASEPTOL (DEKAPURE) Gel is pumped into the spittoon bowl during the tube washing process.
  • Page 46: Reprocessing Kavo Screen One / Kavo Screen Hd

    Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.11 Reprocessing KaVo Screen One / KaVo Screen HD 2.11 Reprocessing KaVo Screen One / KaVo Screen HD Cleaning and disinfection NOTICE Non-colour-fast textiles. Non-colour-fast textiles can cause permanent discoloration.
  • Page 47 Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.12 Reprocessing of the service table 1568 ▶ Clean the surfaces with a soft cloth and water, and wipe-disinfect them with an approved surface disinfectant. Approved disinfectants: ▪...
  • Page 48: Replacing And Checking Cleaning And Germ Reduction Agents

    10 seconds. Note Replenish KaVo OXYGENAL 6 only if there is a beep every 10 seconds. The signal can be turned off as follows: Press the "Intensive germ reduction" but- ton briefly. The next time the treatment centre is turned on, the beep is is- sued again.
  • Page 49: Replenish Oxygenal 6 When Using The Water Bottle

    Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.1 Replenish OXYGENAL 6 with patient part assembly kit / without patient part assembly kit Note Do not fill anything into the central opening (tumbler drain).
  • Page 50 ▶ Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility. ▶ Guide the tube through the orifice of the coarse sieve of the KaVo OXY- GENAL 6 bottle. ▶ Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL 6 bottle.
  • Page 51 In response to over-filling, the excess quantity above the tick mark is aspir- ated automatically to ensure exact dosing. ▶ Place the water bottle in the KaVo OXYGENAL 6 bottle. Make sure that the spout of the KaVo OXYGENAL 6 dosing facility (nose) is situated right above the opening of the water bottle.
  • Page 52: Check The Oxygenal 6 Concentration

    Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.2 Check the OXYGENAL 6 concentration Note Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended periods of non-use after condensation of the water.
  • Page 53 Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.2 Check the OXYGENAL 6 concentration ▶ Immerse a Merckoquant Peroxide Test Strip in the liquid. ð Compare the colours and read the value: 5 mg/l is shown for a concentra- tion of 200ppm.
  • Page 54: Replacing The Dekaseptol Gel Bottle (Dekamat Assembly Kit Only)

    Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.3 Replacing the DEKASEPTOL Gel bottle (DEKAmat assembly kit only) H2O2 concentration Measuring cup Ratio water : treated Value shown on test water strip 0.02 % (200 ppm) 40 ml...
  • Page 55 Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.3 Replacing the DEKASEPTOL Gel bottle (DEKAmat assembly kit only) ▶ Putting the cover back on the unit base. Note Completely empty packages can be disposed of at recycling facilities.