Target group ...................... 4 2 Setup methods according to ISO 17664 ................5 2.1 References........................5 2.2 General instructions for the reprocessing of KaVo treatment units........6 2.2.1 Overview ESTETICA E50 Life reprocessing............. 7 2.3 Reprocessing of the surfaces of treatment unit / upholstery ..........8 2.4 Preparation of the dentist unit ..................
1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. Note Comply with the Instructions for Use of the ESTETICA E50 Life treatment unit! 1.1.1 Abbreviations Explanation brevi- ation...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.1 References 2 Setup methods according to ISO 17664 2.1 References The reprocessing consists of the following basic steps: ▪ Cleaning and disinfection ( manually or automatically) ▪...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units 2.2 General instructions for the reprocessing of KaVo treatment units Please comply with the general reprocessing instructions at all times. Detailed product-specific reprocessing instructions are provided subsequently.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment units Manual drying ▶ Supply dry, clean and uncontaminated compressed air according to ISO 7494-2 only. ▶ Blow off the outside and inside with compressed air until water drops are no longer visible.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.3 Reprocessing of the surfaces of treatment unit / upholstery Reprocessing items/medical Cleaning Cleaning Disinfec- Disinfec- Sterilisa- devices manual automatic tion tion tion manual automatic automatic Assistant element...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Approved disinfectants: ▪ CaviCide / CaviWipes (Metrex) ▪ Microcide AF Liquid (Schülke & Mayr) ▪ FD 322 (Dürr Dental) ▪ Incidin Liquid (Ecolab) ▶...
Page 10
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Cleaning and disinfection of S-table handle and holder pad/anti-slip mat ▶ Press the push button and pull off the handle. ▶ Clean the handle under flowing water.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit 2.4.2 Reprocessing of the instrument holder of the TM table Note The handpiece holder (Mat. no. 1.004.4838) can be removed for better cleaning.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Pull the snap-in buttons outward evenly and carefully remove the instru- ment holder. ▶ Clean the instrument holder under running water.
Page 13
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Attach the germ reduction attachment. with patient part assembly kit / without patient part assembly kit ▶ Take the handpieces off the hoses.
Note Wet the O-rings of the gripping sleeves after sterilisation using silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replacing the cannula, the air channel must be blown out before the start of treatment to ensure that the air exiting during treatment is dry.
Page 15
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Clean the key part, gripping sleeve and cannula by brushing them under running water (at least tap water quality). ▶ Immediately remove contamination from materials used in the dental prac- tice (impression materials, caustic chemicals).
Page 16
Requirement The multifunctional handpiece has been reprocessed. ▶ Apply KaVo silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510) to the O-rings of the joint between the gripping sleeve and cannula. Use cotton buds for this purpose.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit NOTICE Damage caused by over-heating. Damage to the gripping sleeve. ▶ Immediately remove the parts from the steriliser after the sterilisation cycle is completed.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Note The turbine return air filter is in the same position in the dentist element vari- ant with S table. 2.4.7 Reprocessing of the kit for physiological saline solution The kit for physiological saline solution is optional.
Page 19
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Remove the old hose set sterile and dispose of it. ▶ Make sure that the flexing system is horizontal (turn manually if necessary).
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Open the hose clamp ⑦ before startup. Sterilisation kit for physiological saline solution Note The hose set sterile (10 pcs.) (Mat. no. 1.012.4601) is designed for single use and must be changed after each application.
Page 21
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Cleaning Note Disinfect the camera before cleaning. Disinfect after cleaning if necessary. Exterior cleaning by hand NOTICE Damage to the camera due to improper cleaning.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit Approved disinfectants: ▪ MIKROZID Pump Spray - manufacturer: Schülke & Mayr ▪ Dürr System Hygiene FD 322 Spray disinfection - manufacturer: DÜRR Dental GmbH ▪...
Page 23
▶ For programme settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the washer disinfector. ▶ In order to prevent negative effects on the KaVo medical device, make sure that the interior and the exterior of the medical device are dry after comple- tion of the cycle, and then lubricate it immediately with care products from the KaVo care system.
Page 24
For times and concentrations, please refer to the instructions of the manufac- turers of the cleaning/disinfection agent. Manual external disinfection KaVo recommends the following products based on compatibility of the materi- als. The microbiological efficacy must be confirmed by the disinfectant manu- facturer.
Page 25
▶ For programme settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the washer disinfector. ▶ In order to prevent negative effects on the KaVo medical device, make sure that the interior and the exterior of the medical device are dry after comple- tion of the cycle, and then lubricate it immediately with care products from the KaVo care system.
Page 26
Use of worn tips Ultrasound tips are subject to wear and tear and become shorter during use. Worn tips are less effective and might cause pain to the patient. KaVo recommends using the components only for the specified service life. 26 / 58...
Page 27
Sterilisation Sterilisation of handpieces NOTICE Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products. NOTICE Contact corrosion due to moisture. Damage to product. ▶ Immediately remove the product from the steam steriliser after the steril-...
Page 28
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit ▶ Place the cleaned and disinfected handpieces separately in sterilisation packages and weld them to be sealed, (e.g. KaVo STERIclave bags Mat. no. 0.411.9912). 135°C Steriliser with a triple pre-vacuum for at least 4 minutes at 134 °C ±...
▶ Place the cleaned and disinfected tips, endo files, file holders, endo wrenches and torque wrenches separately in sterilisation packages (e.g. KaVo STERIclave bags Mat. no. 0.411.9912) and weld them to be sealed or sterilise them in a sterilisation cassette (e.g. KaVo sterilisation cassette Mat.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit Cleaning and disinfection of the suction hose connector ▶ Clean the disassembled parts under running water. ▶ Thermodisinfect the disassembled parts for at least 10 minutes at 93 °C...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit 2.6.2 Preparation of the tumbler holder and filler NOTICE Water in the device. Material damage from leaking water upon actuation of the "Tumbler filler" or "Bowl flush"...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit The tumbler filler can be subjected to thermodisinfection. Sterilisation of the tumbler holder and filler NOTICE Damage caused by sterilisation. ▶ Do not sterilise the tumbler holder and filler.
(Mat. no. 0.489.3451). ▶ If the Water bottle kit is used with the enclosed dosing attachment (Mat. no. 1.002.0287), add the proper amount of KaVo OXYGENAL 6 (Mat. no. 0.489.3451) with each filling. For the correct amount, please refer to the instructions of the dosing attachment for water germ reduction.
During normal operation of the treatment centre, the continuous germ reduc- tion function secures continuous germ reduction in the water conducting sys- tems due to the automatic dosing of the germ reduction liquid, KaVo OXY- GENAL 6. Rinsing programme and intensive germ reduction facilitate the rinsing and germ reduction of the water-supplying system.
Page 35
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system Place the instruments on the germ reduction attachment with patient part assembly kit / without patient part assembly kit The cannula of the triple-function and multifunctional handpieces must be re- moved for the rinsing program and the intensive germ reduction.
Page 36
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system ▶ Insert the triple-function or multifunctional handpieces without the cannula into the clamping device of the germ reduction attachment. Make sure that the "W"...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system The rinsing program proceeds in four steps. For rinsing programme 01, the cycle starts in step 2, and for rinsing programme 02, the cycle starts in step 4.
Page 38
During intensive germ reduc- tion, the afferent water ducts are automatically filled with a higher concentra- tion of KaVo OXYGENAL 6. The intensive germ reduction program lasts ca. 45 minutes. The increased hydrogen peroxide concentration acts in the system for at least 30 minutes.
Page 39
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.8 Automatic hygiene functions of the water conducting system Step Description Germ reduction of the triple-function and multifunctional handpiece and the spittoon bowl Duration: 60 seconds Germ reduction of the handpieces...
Page 40
The unit is now ready for use. Additional intensive germ reduction during period of non-use KaVo recommends carrying out an additional intensive germ reduction at uni- versities during semester breaks. This flushes the parts of the unit and replaces the germ reduction liquid.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- 2.9 Reprocessing and automatic hygiene functions of the suction system 2.9.1 Reprocessing of the suction system including...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- ter (limited rinsing). The water is aspirated from the Aquamat through the filter housing to the separation system. The spittoon bowl rinsing is alternately turned on and off.
Page 43
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- ð An automatically dosed amount of water is supplied by the vacuum. An acoustical signal is emitted after the process is completed (approx. 3 seconds).
Page 44
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.9 Reprocessing and automatic hygiene functions of the suction sys- Preparation ▶ Fit the hose cleaning attachment (Mat. no. 0.763.8021) to the tap of the Aquamat. E50 Life including CENTRAmat / DEKAmat assembly kit ▶...
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.10 Preparation of the amalgam separator Starting the tube washing function ▶ Press the "HYDROclean" button. ð The tube washing function starts. Note DEKASEPTOL (DEKAPURE) Gel is pumped into the spittoon bowl during the tube washing process.
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.11 Reprocessing KaVo Screen One / KaVo Screen HD 2.11 Reprocessing KaVo Screen One / KaVo Screen HD Cleaning and disinfection NOTICE Non-colour-fast textiles. Non-colour-fast textiles can cause permanent discoloration.
Page 47
Care instructions ESTETICA E50 Life 2 Setup methods according to ISO 17664 | 2.12 Reprocessing of the service table 1568 ▶ Clean the surfaces with a soft cloth and water, and wipe-disinfect them with an approved surface disinfectant. Approved disinfectants: ▪...
10 seconds. Note Replenish KaVo OXYGENAL 6 only if there is a beep every 10 seconds. The signal can be turned off as follows: Press the "Intensive germ reduction" but- ton briefly. The next time the treatment centre is turned on, the beep is is- sued again.
Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.1 Replenish OXYGENAL 6 with patient part assembly kit / without patient part assembly kit Note Do not fill anything into the central opening (tumbler drain).
Page 50
▶ Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility. ▶ Guide the tube through the orifice of the coarse sieve of the KaVo OXY- GENAL 6 bottle. ▶ Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL 6 bottle.
Page 51
In response to over-filling, the excess quantity above the tick mark is aspir- ated automatically to ensure exact dosing. ▶ Place the water bottle in the KaVo OXYGENAL 6 bottle. Make sure that the spout of the KaVo OXYGENAL 6 dosing facility (nose) is situated right above the opening of the water bottle.
Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.2 Check the OXYGENAL 6 concentration Note Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended periods of non-use after condensation of the water.
Page 53
Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.2 Check the OXYGENAL 6 concentration ▶ Immerse a Merckoquant Peroxide Test Strip in the liquid. ð Compare the colours and read the value: 5 mg/l is shown for a concentra- tion of 200ppm.
Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.3 Replacing the DEKASEPTOL Gel bottle (DEKAmat assembly kit only) H2O2 concentration Measuring cup Ratio water : treated Value shown on test water strip 0.02 % (200 ppm) 40 ml...
Page 55
Care instructions ESTETICA E50 Life 3 Replacing and checking cleaning and germ reduction agents | 3.3 Replacing the DEKASEPTOL Gel bottle (DEKAmat assembly kit only) ▶ Putting the cover back on the unit base. Note Completely empty packages can be disposed of at recycling facilities.