1. On the network analyzer select Response > Cal Wizard.
a. If using a mechanical cal kit, select SmartCal (Guided Calibration) > Next.
b. If using an ECal module, connect the ECal to an available PNA USB port and select
Use Electronic Calibration (ECal) > Next.
2. Continue following the Cal Wizard prompts. In the Select Calibration Ports and ECal
Module dialog box press Select All, or select the ports you are calibrating and select
Next.
Figure 26
3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the
Cal Wizard prompts and press Measure after each connection.
4. At the Calibration Completed prompt select Save As User Calset, type the name and
Save. See
5. After calibrating Test Set ports, use a quality load and short to verify the calibration on
each port or end of the test cable. Measure reflection and confirm the return loss is as
expected. If the result is not as expected, repeat the calibration without the Test Set and
ensure that the PNA is in standard (non-multiport) mode.
NOTE
26
This Manual:http://www.manuallib.com/agilent/u3025ae10-test-set-user-guide.html
12-Port Calibration
Figure 32 on page
34.
If measurement errors occur, ensure the newest version of firmware is
installed on the PNA. Measurement errors can be a result of firmware
algorithms. Consult with Agilent Service or firmware web page for the latest
PNA Option 551 firmware revisions and history.
http:\\na.tm.agilent.compna\firmware\firmware.htm.