Transducer Test - HBM MP85A Series Operating Manual

Hide thumbs Also See for MP85A Series:
Table of Contents

Advertisement

Started
Ready
Running
Result valid
Process OK
Fig. 10.2
Process-optimized procedure
This means that result and curve files may not have been transferred if a new
measurement is started immediately. You should therefore only choose this
method if very fast processes have to be monitored and the results and curves
are merely used for random spot checks.

10.3 Transducer test

Initiate the transducer test via a digital input or a bus signal. At low filter limit
frequencies, you have to wait until the filter settling time has elapsed before the
test. A pulse of at least 5 ms duration at the digital input is required for the
transducer test. The result is typically available another 10 ms later.
MP85A
Start (ext. signal "Start"
or condition fulfilled)
Stop (condition fulfilled) or end, if ext. signal for end
condition is used and signal "End" is active
Offline evaluation
Time until stop (follow-up time, only possible when using
Measurement
Initialization (typ. 25 ms)
Reaction time (typ. 2 ms for external signal)
A0293_06_E00_03 HBM: public
Communication with a control system
(internal) end condition fulfilled
(with ext. signal "End" this is
already fulfilled)
Signal is
only
conditions
data is
saved to RAM
Save to MMC/PC
Save (to RAM)
set if
end
are fulfilled
and
internal stop condition)
95