Error messages/operating state (LED display)

Depending on the display mode, various error messages may be displayed on
the LCD display of the MP85A process controller (or the PME Assistant),
instead of the measured value.
Current errors are displayed continuously. Press the
"ERROR" display mode.
Error
message
TransdErr
Input signal overflow, transducer
not connected, transducer
incorrectly connected, amplifier not
adapted to transducer model, no
sense leads connected
ADC Ovfl.
Input signal of measurement
channel A/D converter too large
Grv Ovfl.
Gross value overflow of a
measurement channel
Scal.Err
Input characteristic curve too steep Change input characteristic curve
Flash error
An error occurred while reading out
data from the flash EPROM of the
MP85A process controller.
CAN bus
An error has occurred on the CAN
error
bus
MP85A
Error messages/operating state (LED display)
Cause
A0293_06_E00_03 HBM: public
+
key to go to the
Remedy
Connect transducer, see pin
assignment, connect sense leads
Adapt hardware measuring range
(menu: Preparing for measure­
ment/Amplifier/Transducer)
Reduce display by one decimal
place (menu: Preparing for mea­
surement/Amplifier/Transducer)
(menu: Preparing for measure­
ment/Amplifier/Calibrate
characteristic curve)
This may be a one-off read error,
please repeat the process. If the
error occurs again, please
contact HBM Service.
Check that the termination
resistors are present or whether
a channel is faulty. Then switch
the devices back on. If this does
not help, connect the devices to
the CAN bus one at a time, to
find out which device is faulty.