FAQs: Frequently asked questions (MP85A and PME Assistant)
13.27 What does a tolerance band look like?
Here, the measured curve must remain within the area covered by the toler
ance band, i.e. it must not enter or exit.
13.28 What does an envelope curve look like?
With an envelope curve, the measured curve must enter or exit at the sides.
13.29 What is TEDS?
Transducer electronic data sheet. The TEDS module consists of a chip with a
number that is globally unique (sensor ID), which is generally installed in the
transducer. It contains all the transducer data in accordance with standard
IEEE 1451.4.
Standard IEEE 1451.4 permits various methods of reading out the data from
the TEDS module. In the MP85A, different readout processes are used,
150
Measured curve
Tolerance band
Measured curve
Envelope curve
A0293_06_E00_03 HBM: public