13.12 How long does zero balance take and what does it
involve?
At low filter limit frequencies, you need to wait for the filter settling time to
elapse before zero balance. For zero balance, a pulse of at least 5 ms duration
is required at the digital input. Zero balance is completed another 5 ms later.
Zero balance
Ready
Wait for filter
settling time
Only level changes, i.e. the edges, are analyzed at the inputs.
13.13 What takes place during a transducer test and what are
the important points?
At low filter limit frequencies, you need to wait for the filter settling time to
elapse before the test. A pulse of at least 5 ms duration at the digital input is
required for the transducer test. The result is typically available another 10 ms
later.
134
min. 5 ms
typ. 5 ms