A3.6 Blown Fuses - WALLACH FETAL2EMR 902300 User Manual

Fetal monitor
Table of Contents

Advertisement

fluctuant TOCO baseline
Too
high
TOCO
sensitivity (higher than
100 unit)

A3.6 Blown Fuses

Switch off the monitor and remove the power cord before changing the fuse.
Replace the blown fuse when needed.
The two fuses of the monitor are located on the rear panel; Their specifications are:
Size: Ф5mm*20mm; Model: T2AH250V.
To replace a fuse:
1. Place the monitor on a flat surface and remove the power cord.
2. Pull the fuse container out as far as it can go.
3. Use a screw driver or a pair of pliers to push the fuse up from the bottom of the
container.
The belt has no elasticity.
Maternal movement.
Frequent fetal movements.
The body pressure from uterus to
TOCO transducer is far higher
than the average numeric.
WARNING
- 69 -
Appendix 3: Troubleshooting
Replace the belt.
Request that the patient reduce
movement.
Delay the monitoring.
Confirm good contact to patient
skin with TOCO transducer.
Change the position of TOCO
transducer, if necessary.