all the way into one of the holes on the holder. Make sure that the transducer cable is on the
bottom.
To place the remote event marker, put the small end of the marker into the hole as far as it can go.
Placing the transducer
NOTE:
In the process of monitoring, the transducer that is placed in the holder may be affected and
thereby produce interfering signals. Therefore, when monitoring a patient, it is recommended to
remove or disconnect the transducer that is not in use.

3.8 Connecting Power Cable

 Make sure the AC power supply of the monitor complies with the following specifications:
100V-240V~, 50Hz/60Hz.
 Connect the power cable provided with the monitor to the monitor's power socket. Connect
the other end to a hospital grade grounded 3-wire AC outlet.
If the protective grounding (protective earth) system is in question, the monitor must be
powered only by the battery.
NOTE:
1. Make sure the monitor and the power outlet are placed where it is easy to connect and
disconnect the power cord.
2. When the power is interrupted, the device switches to inner power and operates normally if a
battery is installed. If a battery is not installed, the monitor shuts down and resumes the
previous settings at the subsequent operation.
3. After the AC power supply is connected, please wait for at least 2 seconds before pressing the
POWER switch to turn on the monitor.
Placing the marker
WARNING
- 25 -
Chapter 3: Installation Guide