A1.1 Environmental Specifications

Working
Transport and Storage

A1.2 Physical Specifications

Monitor
Dimensions and Weight
Power Supply
Standards Compliance
Anti-electric Shock Type
Anti-electric Shock Degree
Degree of Protection against Harmful
Ingress of Water for the Transducers
Degree of Safety in Presence of
Flammable Gases
Disinfection/Sterilizing Method
EMC
Temperature:
Relative Humidity:
Atmospheric Pressure:
Temperature:
Relative Humidity:
Atmospheric Pressure:
Size (depth x width x height) :
Weight:
Operating Voltage:
Operating Frequency:
Input Power :
Battery:
UL/IEC/EN 60601-1, IEC/EN 61157, IEC/EN 60601-2-37,
IEC/EN 60601-1-2, IEC 60601-2-49.
Class I equipment with internal power supply
FHR1, FHR2, TOCO, FM, IUP
DECG
IPX 8
Equipment not suitable for use in presence of
flammable gases
Refer to this user manual for details
Group I Class A
- 61 -
Appendix 1: Product Specifications
+5 ºC ~ + 40 ºC ( +41 ºF ~ +104 ºF)
25% ~ 80% (non-condensing)
860hPa ~ 1060hPa
-20 ºC ~ +55 ºC (-4ºF ~ +131 ºF)
25% ~ 93% (non-condensing)
700hPa ~ 1060hPa
350mm x 300mm x 104mm
13.78" x 11.81" x 4.09"
Approx. 3.5 kg/ 6.6 lbs.
100V-240V~
50Hz/60Hz
70VA
14.8V/4200mAh
BF
CF