Appendix 3 Troubleshooting

A3.1 No Display

Error Condition
Power indicator is off.

A3.2 Noise

Error Condition
Noise

A3.3 Recorder Error

Error Condition
Paper jam
Recorder does not work.
Possible Cause
Power cable is loose.
The fuse is blown.
The battery runs out of power.
Possible Cause
Volume set too high
Interference from mobile phone or
other source.
Possible Cause
Paper is loaded incorrectly or
paper is damp.
Wrong paper is being used.
The recorder is not turned on.
Out of paper.
The paper drawer is not locked.
Wrong paper is being used.
- 67 -
Appendix 3: Troubleshooting
Corrective Action
Tighten the power cable.
Change the fuse.
Connect to AC power supply.
Corrective Action
Turn down the volume.
Turn
off
or
remove
interference source.
Move the monitor to a place
with less interference.
Corrective Action
Load paper correctly and keep
paper away from moisture.
Use the paper that is provided
by the manufacturer, or contact
the manufacturer for service.
Press the PRINT key.
Load paper.
Close the drawer cover and
press it until both latches are
locked in position.
Use the paper that is provided
by the manufacturer, or contact
the manufacturer for service.