Cleaning Of Accessories - WALLACH FETAL2EMR 902300 User Manual

Fetal monitor
Table of Contents

Advertisement

Clean the screen with a dry soft cloth. If necessary the screen may be cleaned with a neutral
detergent or isopropyl alcohol. Do not use chemical solvents, acidic or alkali solutions.
1 Although the monitor is chemically resistant to most common hospital cleaners and
non-caustic detergents, different cleaners are not recommended and may stain the
monitor.
2 Many cleansers must be diluted before use. Follow the manufacturer's directions
carefully to avoid damaging the monitor.
3 Do not use strong solvent, for example, acetone.
4 Never use an abrasive such as steel wool or metal polish.
5 Do not allow any liquid to enter the product, and do not immerse any part of the
monitor into liquid.
6 Avoid pouring liquids on the monitor while cleaning.
7 Do not leave any cleaning solution on the surface of the monitor.

10.2.2 Cleaning of Accessories

(1) Cleaning of Transducers
To clean the US transducer and TOCO transducer:
Wipe them with a soft cloth dampened in cleaning solution;
Clean them with a soft cloth dampened in water;
Air-dry them or wipe the remaining moisture with a soft dry cloth.
The recommended cleansers for accessories are listed below:
Accessory
Ultrasound Transducer
TOCO Transducer
1 Do not immerse accessories in liquid. Only clean the outer surface of the connectors,
making sure no liquid goes into the connectors.
2 Be sure the temperature of cleaning solutions does not exceed +45 ºC (+113 ºF).
3 After cleaning accessories, confirm that no remaining cleanser is on the surface.
Chapter 10: Maintenance and Cleaning
CAUTION
Cleansers
BURATON LIQUID
MIKROZID
ETHANOL 70%
SPORACIDIN
CIDEX
CAUTION
- 58 -