Chapter 9 After Monitoring

9.1 Saving Data

The monitor automatically saves the data every five minutes and prior to shutdown, including
fetal monitoring traces and maternal information. The maximum capacity is 12-hour data.
When the monitor is switched on again, the saved data will be loaded. You can review the data or
print it at a high speed.
Turn off the monitor as described in section 9.3 Switching Off. Interrupted power-off may
cause data loss and failure in data reloading.

9.2 Completing the Monitoring

After monitoring is completed,
1. Remove transducers or electrodes from the patient; wipe the remaining gel off the patient
and the transducer with a clean soft cloth or tissue.
2. Tear off the printed recorder paper along the perforation.
NOTE:
1 The remaining coupling gel on the ultrasound transducer may cause the monitor to create
noise. Therefore, the gel should be wiped off the transducer after monitoring.
2 After the fetus is delivered, the monitor may pick up signals of the umbilical cord and display
a trace/numeric value. To avoid misinterpretation, it is recommended the transducers be
removed from the patient and the monitor switched off immediately after the fetus is
delivered.

9.3 Switching Off

1. Press and hold the POWER switch
2. Unplug the power cord.
After the monitor is switched off, please wait for at least 10 seconds before switching it
on again.
CAUTION
for at least 1 second to switch off the monitor.
CAUTION
- 54 -
Chapter 9: After Monitoring