7.4.4 Manual Fetal Movement Monitoring (MFM)

The MFM monitoring result comes from the patient's feeling of fetal movement. The count will
be displayed on the screen in MFM numeric window area.
1. Insert the remote event marker connector into the MARK socket on the monitor.
2. Let the patient hold the remote event marker in hand; ask her to press the key when a fetal
movement is felt. Continuous movements in 5 seconds are considered to be one
movement and the patient should be instructed to press the key only once.

7.4.5 Changing MFM Volume

The monitor gives a sound when the remote marker key is pressed, and the volume is adjustable.
To change the MFM volume,
1. Select the setup key
2. Select Fetus > MFM Volume.
3. Select Low or High (default).
4. Select OK.

7.5 Starting Monitoring

After the START key
transducer connections, clears the MFM count and the monitoring timer. A start symbol
presented on the screen.
If the Auto start printing function is disabled, press the PRINT button
NOTE:
Pressing the START key indicates a new patient is being monitored. Therefore, please remember
to press the START key when monitoring a new patient.

7.6 Inputting Maternal Information (Mat. Info)

7.6.1 Auto ID

After you press the START button
(if Mat. Info inputting function is switched off.) The auto-ID consists of the date and time when
the monitoring starts.
NOTE:
Do not press the START button continuously in a short period of time as the system will not
accept the command to indicate a new patient each time.
on the main interface.
is pressed, the monitor automatically zeroes the pressure, checks
, the system creates an auto-ID for the present patient.
- 43 -
Chapter 7: Fetal Monitoring
to start printing.
is