Technical Information

Output Buffer

Table 27 Output buffer
Offset
Size
Unit Res
Byte
Byte
0
1
1
2
Hz
3
2
Hz
252
Range
B
= Start/Stop
7
0 = Stop, 1 = Start
B
= Low Speed
6
This bit set the pump target speed to "low speed value"
(Output Buffer, Offset 3), otherwise the target speed is set
to "high speed value" (Output Buffer, Offset 1).
0 =Off, 1 =On
B
– Soft Start
5
This bit activates the soft start function which run-up the pump
smoothly. Please activate this function only if the pump has
been stopped for at least 3 months.
0 = On, 1= Off
B
= Vent Valve Command
4
0 = Close, 1 = Open
Valid only if in Par Buff, Offset-20, B
B
= Purge Valve Command
3
0 = Close, 1 = Open
Valid only if Par Buff, Offset-20, B
B
– Not used
2
B
= Operating Mode
0-1
These bits define from which source the controller has to get
commands (readings are always available). In order to operate
the controller by Profibus these bits must be set to 3.
0 = Remote I/O
1 = Serial
2 = Front panel
3 = Profibus
1100 -
High Speed Value
1350
Set the high rotational speed value
1100 -
Low Speed Value
1350
Set the low rotational speed value