Technical Information
Where:
V
is the gauge output voltage;
g
V
out
V
g_min
from the gauge technical specifications or from the following table.
Table 12
ID
1
2
3
4
6
9
10
11
12
Agilent TwisTorr 305 FS Remote Controller User Manual
is the Analog Output voltage;
V
is the gauge voltage measuring range. These values can be drawn
÷
g_max
Gauge Model
CT-100
FRG-700/702
IMG-500
PVG-500/502
PCG-750/752
CDG-500 T1000
CDG-500 T0100
CDG-500 T0010
CDG-500 T0001
Pressure range [mbar]
1.33E-2 to 1.01E3
5E-9 to 1E3
2E-9 to 1E-2
5E-4 to 1E3
5E-5 to 1.5E3
2.66E+1 to 1.33E3
2.66E+0 to 1.33E2
2.66E-1 to 1.33E1
2.66E-2 to 1.33
Vg-min ÷
Vg-max [V]
1.11 ÷ 8.25
1.82 ÷ 8.6
1.8 ÷ 8.5
1.9 ÷ 10
0.61 ÷ 10.23
0.2 ÷ 10
0.2 ÷ 10
0.2 ÷ 10
0.2 ÷ 10