Agilent Technologies TwisTorr 305 FS User Manual page 239

Hide thumbs Also See for TwisTorr 305 FS:
Table of Contents

Advertisement

Technical Information
R1 (R2) TYPE: selects the type of reference measurement to drive the relay.
Possible types:
R1 (R2) SET VALUE: Sets the reference value at which the relay R1 (R2) is
activated. This parameter has no effect if NORMAL (previous parameter) is
selected as SETPOINT TYPE.
R1 (R2) HISTERESYS: Sets, in percentage, the hysteresis of the relay
R1 (R2) MASK TIME: sets the masked time (in hh:mm:ss), measured from start,
within which the SETPOINT R1 (R2) is not driven even if the threshold value has
been reached. This parameter is not available if SETPOINT TYPE is set to
NORMAL
R1 (R2) OUTPUT LOGIC: Selects the control logic of the SETPOINT output.
Agilent TwisTorr 305 FS Remote Controller User Manual
Frequency: driving frequency of the pump
Power: power supplied to the pump
Time: time from start
Normal: achievement of normal condition
Pressure: pressure read by the gauge, if present
positive logic (the relay is energized when the measurement exceeds the
reference)
negative logic (the relay is energized when the measurement is below the
reference).
239