Technical Information

Input/Output Menu

This menu contains parameters that determine the operating mode of the inputs
and outputs of the controller.
Figure 67
START MODE: Permits selection of one of the three types of functioning of the
START/STOP and INTERLOCK inputs:
Agilent TwisTorr 305 FS Remote Controller User Manual
START/INTERLOCK: The pump is started if the START/STOP input is
activated and is stopped if the START/STOP input is de-activated.
The interlock input must be constantly active (INTERLOCK TYPE
parameter = continuous) or may be active only at the time of starting
(INTERLOCK TYPE parameter = impulsive).