Technical Information

Window Meanings

Table 23
WIN# Read/
Type
Write
0
R/W
L
1
R/W
L
8
R/W
A/L
100
R/W
L
101
R/W
N
102
R/W
N
103
R/W
N
104
R/W
L
105
R/W
N
107
R/W
L
108
R/W
N
109
W
L
110
R/W
L
Agilent TwisTorr 305 FS Remote Controller User Manual
Meaning
Start/Stop (in remote/ Front mode
the window is a read only)
Low Speed Activation
Serial/Remote/Front/FieldBus
Contr. Configuration
Enable/Disable Soft Start
(write only in Stop condition)
R1 Set Point type
R1 Set Point value
(expressed in Hz/W/s/mbar)
R1 Set Point delay: time between the
pump start and the set point check
(seconds)
R1 Set Point signal activation type: the
signal can be "high level active" or "low
level active"; "high level active" indicates
the output is ON when the signal is
above the threshold
R1 Set point hysteresis
(in % of R1 value)
Active Stop (write only in stop)
Enable/disable the stop with active
braking
Serial comm. baud rate
Pump life/Cycle time/Cycle number
reset
Interlock type.
The interlock can be checked at start
(impulse) or continuously
Admitted
Values
Start = 1
Stop = 0
No = 0
Yes = 1
0 = Serial
1 = Remote
2 = Front
3 = Profibus
Enable = 1
Disable = 0
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 = Pressure (available only if the
gauge is connected)
0¸999999
0¸999999
0 = high level active
1 = low level active
0¸100
Enable = 1
Disable = 0
0 = 600 bit/s
1 = 1200 bit/s
2 = 2400 bit/s
3 = 4800 bit/s
4 = 9600 bit/s
To reset write '1'
Impulse = 0
Continuous = 1
Factory
Set
0
0
1
"1" at the first
START, then "0"
3
867
0
0
2
0
4
213