When the Request is "J" (Numerical readings), the Answer will contain a string of 5
characters with the following parameters:
Table 19
Bytes
1
2
3
4
5
When the Request is "K" (Counters readings), the Answer will contain a string of 11
characters with the following parameters:
Table 20
Bytes
1-4
5-8
9-10
11
On request a sample program in QBasic language is available by Agilent.
206
Meaning
Current (0-255 scaled)
Voltage (0-255 scaled)
Rotational speed KRPM
Pump temperature °C (0-254 temperature reading, 255 = fail)
CRC
Meaning
Cycle time
Pump life
Cycle number
CRC