Agilent Technologies TwisTorr 305 FS User Manual page 172

Hide thumbs Also See for TwisTorr 305 FS:
Table of Contents

Advertisement

Instructions for Use
Table 7
Error Messages
SYSTEM OVERRIDE
OVERVOLTAGE
RUN UP TIME
172
The pump is stopped by an emergency stop
signal provided via a remote contact.
Controller power supply circuitry is faulty, or
the Controller received a spike.
Run Up Time out.
Agilent TwisTorr 305 FS Remote Controller User Manual
Remove the controller power cable and
check the emergency condition. Then
reconnect the power cable and press the
START push-button twice to start the
pump.
Press the START push-button twice to start
the pump.
Should the message still be present, call
the Agilent service.
Check for the Run Up Time set-up. It has to
be adjusted accordingly to the customer
process. Check the amount of the gas load
which is flowing inside the pump.