TwisTorr 305 FS Remote Controller Models X3506-64130, X3506-64131 Manuale di istruzioni Bedienungshandbuch Notice de mode d’emploi Manual de istrucciones 用户手册 ユーザーマニュアル User Manual 87-901-059-01 01/2020...
Page 2
Do not “Restricted computer software” as proceed beyond a WARNING notice defined in FAR 52.227-19 (June 1987) until the indicated conditions are or any equivalent agency regulation or fully understood and met. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Istruzioni per l’uso Informazioni Generali Immagazzinamento Preparazione per l’installazione Installazione Procedure di uso Manutenzione Smaltimento Messaggi di errore Gebrauchsanleitung Allgemeines Lagerung Vor der Installation Installation Gebrauch Bedienung Wartung Entsorgung Fehlermeldungen TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 5/190...
Page 6
Preparation pour l‘installation Installation Utilisation Procedures d'utilisation Entretien Mise au rebut Messages d'erreur Manual de istrucciones Información general Almacenamiento Preparación para la instalación Instalación Procedimientos de uso Mantenimiento Eliminación Mensajes de error 6/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 8
TwisTorr 305 FS Remote Controller Description Controller Specifications Controller Outline Controller Connection R1 – R2 Output Diagrams Programmable Analog Output Diagrams J2 – Serial Connector Vacuum Link App Letter Protocol Description Examples 8/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 9
Contents How to Use by Front Panel Programming How to Use by Remote I/0 How to Use in Serial Mode Profibus Option Error Messages Orderable Parts TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 9/190...
Page 10
Contents Page intentionally left blank 10/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
TwisTorr 305 FS Remote Controller User Manual Istruzioni per l’uso Informazioni Generali Immagazzinamento Preparazione per l’installazione Installazione Comandi, Indicatori e Connettori Descrizione pannello frontale Procedure di uso Accensione del Controller Avvio della Pompa Arresto della Pompa Manutenzione Smaltimento Messaggi di errore...
I controller della serie TwisTorr 305 FS Remote sono dei convertitori di frequenza, controllati da un microprocessore, realizzati con componenti a stato solido e con capacità di autodiagnostica e autoprotezione.
Durante il trasporto e l’immagazzinamento dei controller devono essere soddisfatte le seguenti condizioni ambientali: temperatura: da -20 °C a +70 °C ▪ umidità relativa: 0 – 95 % (non condensante) ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 13/190...
Non disperdere l'imballo nell'ambiente. Il materiale è completamente riciclabile e risponde alla direttiva CEE 85/399 per la tutela dell'ambiente. Figura 1 Imballo dei controller 14/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
5 °C a +45 °C; ▪ umidità relativa: 0 – 95 % (non condensante). ▪ Per gli altri collegamenti e I'installazione degli accessori opzionali, vedere la sezione "Technical Information". TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 15/190...
Il connettore di richiusura J1 deve essere lasciato collegato con il suo ponticello NOTA se non viene effettuato alcun collegamento esterno. La pompa di pre-vuoto e la pompa TwisTorr 305 FS possono essere accese contemporaneamente. 16/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 17
Display alfanumerico a cristalli liquidi: matrice di punti, 2 linee x 16 caratteri. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 17/190...
Page 18
Modulo di ingresso alimentazione controller che comprende i fusibili, la presa di alimentazione ed il filtro ENC. Switch a rotazione per impostare l’indirizzo del Profibus. Profibus (opzionale). Connettore seriale per controllo da remoto. Pompa. 18/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Per arrestare la pompa occorre premere il pulsante STOP del pannello frontale. Manutenzione I controller della serie TwisTorr 305 FS Remote non richiedono alcuna manutenzione. Qualsiasi intervento deve essere eseguito da personale autorizzato. In caso di guasto è possibile usufruire del servizio di riparazione Agilent o del "Agilent advanced exchange service", che permette di ottenere un...
70 °C. temperatura ritorni al di sotto della soglia. Premere Oppure. due volte il pulsante START La temperatura del radiatore del per riavviare la pompa. controller è superiore a 60 °C. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 21/190...
Page 22
Agilent per la manutenzione. RUN UP TIME Tempo di Run Up scaduto. Verificare la corretta impostazione del tempo di Run Up in funzione dell’applicazione. Verificare il carico di gas applicato alla pompa. 22/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
TwisTorr 305 FS Remote Controller User Manual Gebrauchsanleitung Allgemeines Lagerung Vor der Installation Installation Gebrauch Steuerungen, Anzeigen und Verbinder des Controllers 28 Fronttafel der Controller Bedienung Einschalten des Controllers Pumpenstart Pumpenstopp Wartung Entsorgung Fehlermeldungen Übersetzung der Originalanleitungen 23/190...
Eingriffen und Mißachtung der einheimischen, hier zur Geltung kommenden Bestimmungen übernimmt die Firma Agilent keinerlei Haftung. Die Controller der Serie TwisTorr 305 FS Remote sind mikroprozessorgesteuerte Frequenzwandler. Sie sind mit Festkörperbauteilen gefertigt und verfügen über ein Selbstdiagnose- und ein Selbstschutzsystem.
Beim Transport und bei der Lagerung der Controller müssen folgende klimatische Verhältnisse eingehalten werden: Temperatur: von -20 °C bis +70 °C ▪ Relative Luftfeuchtigkeit: 0 – 95 % (nicht kondensierend) ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 25/190...
Beim Auspacken vorsichtig vorgehen, damit der Controller nicht fällt oder Stößen ausgesetzt wird. Das Verpackungsmaterial muß korrekt entsorgt werden. Es ist vollständig recyclebar und entspricht der EG-Richtlinie 85/399 für Umweltschutz. Abbildung 1 Verpackung der Controller 26/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Gasen ausgesetzt sind und in denen Explosions- und erhöhte Brandgefahr besteht. Beim Betrieb müssen folgende Umgebungsbedingungen eingehalten werden: Temperatur: von +5 °C bis +45 °C ▪ Relative Luftfeuchtigkeit: 0 – 95 % (nicht kondensierend). ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 27/190...
Schließflansch abgedeckt ist. Der Wiederverschließ-Verbinder J1 muß mit seiner Brücke angeschlossen HINWEIS bleiben, wenn kein externer Anschluß erfolgt. Die Vorvakuumpumpe und die TwisTorr 305 FS Pumpe können gleichzeitig eingeschaltet werden. 28/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 29
TwisTorr 305 FS Remote TwisTorr 305 FS AG Abbildung 2 Fronttafel der Controller TwisTorr 305 FS Remote Taste für die Anzeige der Parameter "cycle number", "cycle time" und "pump life". Taste für die Einstellung des Modus LOW SPEED. Sie ist nur aktiv, wenn der Steuermodus auf Fronttafel eingestellt ist.
Page 30
Gebrauchsanleitung Gebrauch Abbildung 3 Rücktafel der Controller TwisTorr 305 FS Remote Eingangsverbinder der logischen Signale (der Kupplungsverbinder wird mit einer Wiederverschließbrücke geliefert). Stecker für logische Signale und Pumpenmonitor plus programmierbarer Analogausgang. Stromausgangsstecker (24 V= für Pumpenkühlventilator, Vent Valve und Messinstrument).
Zum Stoppen der Pumpe muß die STOP-Taste an der Fronttafel gedrückt werden. Wartung Die Controller der Serie TwisTorr 305 FS Remote sind wartungsfrei. Eventuell erforderliche Eingriffe müssen von dazu befugtem Fachpersonal ausgeführt werden. Bei einem Defekt kann der Agilent-Reparaturdienst bzw. der "Agilent advanced exchange service"...
Der Endabnehmer sollte daher den Lieferanten des Geräts - d.h. die Muttergesellschaft oder den Wiederverkäufer - kontaktieren, um den Entsorgungsprozess zu starten, nachdem er die Verkaufsbedingungen geprüft hat. Weitere Informationen finden Sie unter: http://www.agilent.com/environment/product/index.shtml 32/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Warten bis die Temperatur OVERTEMP. 70 °C überschritten. unter den Schwellenwert Oder gesunken ist. Die Pumpe Die Temperatur des durch zweimalige Kontrollerkühlkörpers ist über 60 °C. Betätigung der START- Taste neustarten. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 33/190...
Page 34
Wartungs-dienst gerufen werden. RUN UP TIME Anlaufzeitüberschreitung. Prüfen Sie die Einrichtung der Anlaufzeit. Sie muss entsprechend dem Kundenprozess angepasst werden. Prüfen Sie die Menge der Gaslast, die in der Pumpe fließt. 34/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
TwisTorr 305 FS Remote Controller User Manual Mode d’emploi Indications generales Emmagasinage Preparation pour l‘installation Installation Utilisation Commandes, Indicateurs et Connecteurs Description du Tableau avant Procedures d'utilisation Allumage du Contrôleur Mise en marche de la Pompe Arrêt de la Pompe...
Les contrôleurs de la série TwisTorr 305 FS Remote sont des convertisseurs de fréquence, contrôlés par un microprocesseur, réalisés avec des éléments à...
Pendant le transport et l'emmagasinage des contrôleurs, il faudra veiller à respecter les conditions environnementales suivantes: température: de - 20 °C à + 70 °C ▪ humidité relative: de 0% à 95 % (non condensante). ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 37/190...
Ne pas jeter l'emballage dans la nature. Le matériel est entièrement recyclable et il est conforme aux directives CEE 85/399 en matière de protection de l'environnement. Figure 1 Emballage du Contrôleur 38/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
+5 °C à +45 °C ▪ humidité relative: de 0 % à 95 % (non condensante). ▪ Pour les autres connexions et pour l'installation des accessoires en option, voir la section "Technical Information". TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 39/190...
Laisser le connecteur de réenclenchement J1 connecté à sa barrette s'il n'est NOTE procédé à aucune connexion extérieure. La pompe à pré-vide et la pompe TwisTorr 305 FS peuvent être mises en marche simultanément. 40/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 41
à zéro du contrôleur et une deuxième fois pour remettre la pompe en marche. Ecran alphanumérique à cristaux liquides: matrice de points, 2 lignes x 16 caractères. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 41/190...
Page 42
Module d’entrée alimentation contrôleur qui comprend les fusibles, la prise d’alimentation et le filtre ENC. Interrupteur à rotation pour établir l’adresse du Profibus. Profibus (facultatif). Connecteur sériel pour contrôle à distance Connecteur pompe. 42/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Pour arrêter la pompe, presser l'interrupteur STOP du tableau frontal. Entretien Les contrôleurs de la série TwisTorr 305 FS Remote n'exigent aucun entretien. Toute opération doit être effectuée par un personnel agréé. En cas de panne, il est possible de s'adresser au Service de réparation Agilent ou bien au "Agilent advance exchange service"...
œuvre le processus de collecte et mise au rebut. Pour en savoir plus, consulter : http://www.agilent.com/environment/product/index.shtml 44/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
70 °C. retourne au-dessous du Ou bien seuil. Presser deux fois La température du radiateur du l'interrupteur START pour contrôleur est supérieure à 60 °C. remettre la pompe en marche. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 45/190...
Page 46
Vérifiez la configuration du temps de démarrage. Elle doit être ajustée en fonction du processus du client. Vérifiez la quantité de charge de gaz qui s’écoule à l’intérieur de la pompe. 46/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
TwisTorr 305 FS Remote Controller User Manual Manual de istrucciones Información general Almacenamiento Preparación para la instalación Instalación Mandos, Indicadores y Conectores Descripción del panel frontal Procedimientos de uso Encendido del controler Puesta en marcha de la Bomba Parada de la Bomba Mantenimiento Eliminación...
Los controlers de la serie TwisTorr 305 FS Remote son convertidores de frecuencia, controlados por un microprocesador, realizados con componentes en estado sólido y con capacidad de autodiagnosis...
Durante el transporte y el almacenamiento de los controlers se deberá cumplir con las condiciones ambientales siguientes: temperatura: de -20 °C a +70 °C ▪ humedad relativa: 0 – 95 % (no condensadora) ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 49/190...
No dispersar el embalaje en el medio ambiente. El material es totalmente reciclable y cumple con la directiva CEE 85/399 para la preservación del medio ambiente. Figura 1 Embalaje de los Controlers 50/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
+5 °C a +45 °C ▪ humedad relativa: 0 – 95 % (no condensadora). ▪ Para otras conexiones y la instalación de los accesorios opcionales, véase la sección “Technical Information”. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 51/190...
El conector di cierre J1 ha de dejarse conectado con su conector puente NOTA si no se efectúa ninguna conexión exterior. La bomba pre-vacío y la bomba TwisTorr 305 FS pueden encenderse simultáneamente. 52/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 53
Display alfanumérico de cristales líquidos: matriz de puntos, 2 líneas x 16 caracteres. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 53/190...
Page 54
Módulo de entrada alimentación controler con fusibles, toma de alimentación y filtro ENC. Conmutador de rotación para configurar la dirección del Profibus. Profibus (opcional). Conector serial para control remoto. Conector bomba. 54/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Para detener la bomba hay que apretar el pulsador STOP del panel frontal. Mantenimiento Los controlers de la serie TwisTorr 305 FS Remote no necesitan ningún mantenimiento. Cualquier operación ha de ser efectuada por personal autorizado. En caso de avería es posible utilizar el servicio de reparación Agilent o del “Agilent advance exchange service”, que permite obtener un controler...
O bien debajo del umbral. Apretar La temperatura del radiador del dos veces el pulsador controler es superior a 60 °C. START para volver a poner en marcha la bomba. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 57/190...
Page 58
RUN UP TIME Tiempo de arranque agotado. del tiempo de arranque. Debe ajustarse según el proceso del cliente. Comprobar la cantidad de carga de gas que fluye dentro de la bomba. 58/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
TwisTorr 305 FS Remote Controller User Manual Instructions for Use General Information Storage Preparation for Installation Installation Controls, Indicators and Connectors Front Panel Description Use Procedure Controller Startup Starting the Pump Pump Shutdown Maintenance Disposal Error Messages Original Instructions 83/190...
The TwisTorr 305 FS Remote Controllers are micro-processor-controlled, solid-state, frequency converter with self-diagnostic and self-protection features. Controller features: Front / Remote / Serial Operation ▪...
When transporting and storing the controllers, the following environmental requirements should be satisfied: temperature: from -20 °C to + 70 °C ▪ relative humidity: 0 – 95 % (without condensation) ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 85/190...
Do not dispose of the packing materials in an unauthorized manner. The material is 100 % recyclable and complies with EEC Directive 85/399. Figure 1 Controller packing 86/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
CE requirements. High voltage developed in the controller can cause severe injury or death. Before servicing the unit, disconnect the input power cable. TwisTorr 305 FS Remote Controller can be used as a bench unit or a rack module, NOTE but it must be positioned so that free air can flow through the holes.
To avoid injury to personnel and damage to the equipment, if the pump is laying on a table make sure it is steady. Never operate the TwisTorr 305 FS pump if the pump inlet is not connected to the system or blanked off.
Page 89
If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump. LCD back-lighted alphanumeric display: dot matrix 2 lines x 16 characters. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 89/190...
Page 90
Power output connector (24 Vdc for pump cooling fan, Vent Valve and Gauge). Controller power entry module consisting of mains fuses, mains socket and EMC filter. Rotary switches for Profibus address setting. Profibus (Optional). Remote control serial connector. Pump cable (5m long). 90/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
To shutdown the pump press the STOP push-button on the controller front panel. Maintenance The TwisTorr 305 FS Remote series controller does not require any maintenance. Any work performed on the controller must be carried out by authorized personnel. When a fault has occurred it is possible to use the Agilent repair service.
Parent Company or a retailer, to initiate the collection and disposal process after checking the contractual terms and conditions of sale. For more information refer to: http://www.agilent.com/environment/product/index.shtml 92/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Wait until the temperature OVERTEMP. temperature exceeds 70 °C. decrease below threshold value. Press the START push- The controller’s radiator button twice to start the temperature is above 60 °C. pump. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 93/190...
Page 94
Check for the Run Up Time set-up. It has to be adjusted accordingly to the customer process. Check the amount of the gas load which is flowing inside the pump. 94/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 96
How to Use in Serial Mode Profibus Option General Installation Configuration Software Operation and State Machine Message Mapping Parameter Buffer Output Buffer Input Buffer Diagnostic Management Diagnostic Buffer Error Messages Orderable Parts 96/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
(ref. TwisTorr 305 FS Remote Controller). Serial protocol is in most of the cases compatible with the previous version (ref. TwisTorr 305 FS Remote Controller) but, due to the additional features that have been implemented on this unit, some serial windows have been revised in the data format/content.
Page 98
TO POWER MAINS FUSE TYPE T 4A SUPPLY VENT GAUGE MAINS CABLE GAUGE CABLE (1 m) GAUGE EXTENTION CABLE (5 m) GAUGE Figure 4 TwisTorr 305 FS Remote Controller – Connections 98/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
The Controller must be positioned in a way that the mains cable can be easily NOTE disconnected from the controller mains power inlet (the controller is not equipped with a mains switch). TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 99/190...
The outline dimensions for the TwisTorr 305 FS Remote Controllers are shown in the following figure. TwisTorr 305 FS Remote Controller can be used as a bench unit or as a rack NOTE module, but it must be positioned so that free air can flow through the holes.
This is a dedicated output for the optional TwisTor 305 pump cooling fan (see orderable parts table). If you already have fan installed, use the available adapter cable to connect to the pump. Figure 6 Fan Connector TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 101/190...
Page 102
Vent Valve CLOSED OPEN Vent Valve Relay Vent Valve Open RAMP STOP NORMAL Win. 126 Time Win. 147 START COMMAND STOP COMMAND Figure 8 Vent Valve diagram in “Auto” mode 102/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 103
If the contact is open, it allows the Imin (on) = 2mA pump running; if the contact is closed, it stops the pump and all the others connected devices (Fan, Vent). TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 103/190...
Page 104
The following figure shows a typical contact logic input connection and the related simplified circuit of the controller. Figure 10 Typical logic input connection J1 – Output Figure 11 Logic output connector 104/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 105
LED, a computer, etc., and the related simplified circuit of the controller. +24 V Minimum 1 kΩ Figure 12 Digital Output Connection Example Minimum 100 kΩ Figure 13 Analog Output Connection Example TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 105/190...
Page 106
Pressure. See following Imax = 60mA measuring voltage figures for more details. All Pin-7 negative the setting can be done by: Pin-14 positive front panel or serial line (win.171 to 176). 106/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 107
When the analog output is set to provide pressure signal, the voltage level is proportional to pressure level and it is generated by control unit elaborating the independent input by the active gauge. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 107/190...
R1 – R2 Output Diagrams R1 – R2 Output Diagrams Figure 14 Diagram 1 – R1 or R2 related to frequency Figure 15 Diagram 2 – R1 or R2 related to power 108/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 109
Diagram 3 – R1 or R2 related to time Figure 17 Diagram 4 – R1 or R2 related to Normal status Figure 18 Diagram 5 – R1 or R2 related to pressure reading (if gauge connected) TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 109/190...
Figure 20 Diagram 2 – Programmable analog output related to power V (Volt) Output Voltage Pump Temp (°C) Figure 21 Diagram 3 – Programmable analog output related to pump temperature 110/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 111
B-C-D = 75 Vac 3-phase output to pump motor stator ▪ E = EARTH PROTECTION ▪ Figure 22 Controller-to-Pump Connector J5 – Gauge Connector This connector is available to connect the Agilent active gauges. Figure 23 Gauge Connector TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 111/190...
Page 112
0 to 10V. The conversion formula is the following: 112/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 113
2.66E+1 to 1.33E3 0.2 ÷ 10 CDG-500 T0100 2.66E+0 to 1.33E2 0.2 ÷ 10 CDG-500 T0010 2.66E-1 to 1.33E1 0.2 ÷ 10 CDG-500 T0001 2.66E-2 to 1.33 0.2 ÷ 10 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 113/190...
Page 114
(win 101) is selected as “pressure” the gauge is always on. The windows from 136 to 140 manage an additional set point (set point 3) dedicated to turn ON or OFF the gauge (win 267=4) 114/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 115
This function can be activated by Front Panel or serial line (Win 107, 167). Active Stop brakes the pump using the motor. The Stop Speed reading reads the pump speed after a stop command. Figure 25 Diagram 5 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 115/190...
Apple Store. The app is available for both Android and iOS. Once installed, you can establish the connection. Using this app you will be able to read and write all controller parameters. 116/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 118
(old system) “letter” ▪ “Window” protocol (new system) ▪ These two protocols can be used as well with 232 or 485 media. Please use “Window” protocol for new development. NOTE 118/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
"C" = Low Speed ON ▪ "D" = Low Speed OFF ▪ "E" = Request for operational parameters ▪ "F" = Pump times zeroing ▪ "G" = Parameters reading ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 119/190...
Page 120
START command "A" in ASCII code = 41 inverted it will be: FF + 1 - 41 = BF. Tab. 6 CHARACTER ASCII CONVERTED ASCII FOR CRC CALCULATION "A" "B" "C" "D" "E" "F" "G" "H" "I " "J" "K" 120/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 121
Answer = after a request from the host, the Controller will answer in one of the following ways: ▪ NACK ▪ Message ▪ When the Request is "A, "B”, "C", "D", "F" the Controller will Answer the ACK or NACK. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 121/190...
Page 122
*The values for current and voltage are given as numbers, scaled from 0 to 255, where 0 corresponds to 0 V and 255 to the full scale voltage (130 V) or current (2,5 A). 122/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 123
Soft Start mode (0 = NO 1=YES) When the configuration parameters have to be changed, send a Request string with "H" characters of the following type: "H" + ..data ..+ CRC TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 123/190...
Page 124
= NORMAL OPERATION 5 = HIGH LOAD = FAILURE 7 = APPROACHING LOW SPEED R2 status 0 = OFF 1 = ON R1 status 0 = OFF 1 = ON 124/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 125
11 characters with the following parameters: Tab. 12 BYTES MEANING Cycle time Pump life 9-10 Cycle number On request a sample program in QBasic language is available by Agilent. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 125/190...
Page 126
The MESSAGE is a string with the following format: <STX>+<ADDR>+<WIN>+<COM>+<DATA>+<ETX>+<CRC> Where: When a data is indicated between two quotes (‘...’) it means that the indicated NOTE data is the corresponding ASCII character. 126/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 127
Using the RS 485 interface, the message structure remains identical to the one NOTE used for the RS 232 interface, the only difference being that the value assigned to the ADDRESS <ADDR> The controller can answers with the following response types: TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 127/190...
Page 128
Win Disabled 1 byte (0x35) the specified window is Read Only or temporarily disabled (for example you can’t write the Soft Start when the Pump is running) 128/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Technical Information Examples Examples Command: START Source: PC Destination: Controller ADDR WINDOW Source: Controller Destination: PC ADDR Command: STOP Source: PC Destination: Controller ADDR WINDOW Source: Controller Destination: PC ADDR TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 129/190...
Page 130
Technical Information Examples Command: SOFT-START (ON) Source: PC Destination: Controller ADDR WINDOW Source: Controller Destination: PC ADDR Command: SOFT-START (OFF) Source: PC Destination: Controller ADDR WINDOW Source: Controller Destination: PC ADDR 130/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 131
Source: Controller (with address = 3 in stop status) Destination: PC WINDOW DATA (STATUS) Command: REAL SERIAL TYPE Source: PC Destination: Controller (with address = 3 in 485 mode) ADDR WINDOW Source: Controller Destination: PC ADDR WINDOW DATA TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 131/190...
Page 132
4 = 9600 bit/s Pump life/Cycle time/Cycle number reset To reset write ‘1’ Interlock type. Impulse = 0 The interlock can be checked at start Continuous = 1 (impulse) or continuously 132/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 133
Vent valve opening time (1unit = 0.2s; 0¸65535 (from 0 to 0 = infinite) 13107s) (Ref. to “vent connector” paragraph) Actual maximum power limit (W) (It reads the maximun power deliverable to the pump) TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 133/190...
Page 134
Threshold value is meaningful if win101 = 4 Start Output Config. 0 = Starting (Output ON only with pump Status = Starting) 1 = Running (Output ON when the pump is running) 134/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 135
Controller cooling air temperature (°C) Pressure reading Format = X.X E XX Rotation Frequency (rpm) Gauge Status 4 = RidUnknown 3 = GaugeOverRange 2 = GaugeUnderRange 1 = GaugeConnected 0 = NoGaugeConnected TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 135/190...
Page 136
Run Up time control On = 1 Off = 0 TOO HIGH LOAD NO CONNECTION SHORT CIRCUIT PUMP OVERTEMP. OVERVOLTAGE CONTROLL. OVERTEMP. OVERRIDE RUN UP TIME Figure 29 Window N. 206 Bit Description 136/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
“FRONT”; otherwise an error message is returned. If it is set to “REMOTE” mode, the COMMAND LOCKED IN REMOTE error message is displayed while if it is in “SERIAL” mode the “COMMAND LOCKED IN SERIAL MODE” error message is returned. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 137/190...
Page 138
Pressing a third time, the following message Is displayed: Model Number of the controller Serial Number On pressing the key again, the main screen page is redisplayed. Measures ▪ Pressing once, the following is obtained: ° 138/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 139
XXXHz: see description “Pressing once” Measures key I=X.XXA: indicates the current absorbed by the pump V= XXX V: indicates the pump driving voltage Pressing 3 times, the following is obtained: TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 139/190...
Page 140
Other possible indications: Where: GAUGE_ERROR: gauge not connected or broken Where: PRESS UNDER: indicates that the pressure read is below the minimum measurement threshold set (5 E-9 mBar) 140/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 141
RS485 [XX]: the controller accepts commands from the serial port in RS485 mode with address XX (XX from 0 to 31) PROFIBUS [YYY]: the controller accepts commands from the PROFIBUS port with address YYYY TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 141/190...
Page 142
LS → the controller is at LOW SPEED AG → the Gauge is connected to the controller R1 → relay (set point) R1 is active R2 → relay (set point) R2 is active Figure 31 142/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
To avoid injury to personnel and damage to the equipment, if the pump is laying on a table make sure it is steady. Never operate the TwisTorr 305 FS pump if the pump inlet is not connected to the system or blanked off.
Page 144
Pressing the START button, the pump is started and the ▪ display shows the following screen page. VACUUM TECHNOLOGIES COUNTERS MEASURES increase next menu previous decrease LOW SPEED START/STOP TwisTorr 305 FS Remote Turbo-V 301-AG Figure 32 144/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 145
If the LOW SPEED key is pressed, the above screen page changes to: VACUUM TECHNOLOGIES COUNTERS MEASURES increase next menu previous decrease LOW SPEED START/STOP TwisTorr 305 FS Remote Figure 33 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 145/190...
Page 146
Press the MEASURES push button: the display shows: ▪ ▪ ° VACUUM TECHNOLOGIES COUNTERS MEASURES increase next menu previous decrease LOW SPEED START/STOP TwisTorr 305 FS Remote Figure 34 146/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 147
°C = is the temperature of the outer ring of the upper bearing (range 00 to 99 °C) Pressing MEASURES key twice, the following screen page is ▪ displayed: TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 147/190...
Page 148
COUNTERS MEASURES increase next menu previous decrease LOW SPEED START/STOP TwisTorr 305 FS Remote Figure 35 Where: mBar shows the pressure value measured XXXX shows the kind of gauge connected 148/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 149
TwisTorr 305 FS Remote Figure 36 Where: CYCLE XXXX Number of start/stop cycles TIME XXXXXm Time of last cycle (minutes) Pressing again: Where: PUMP LIFE XXXXXX total operating time (hours)of the pump TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 149/190...
The “MEASURES and START/STOP” buttons become “NEXT and ▪ PREVIOUS” and make it possible to select the next parameter (or next digit of a numeric parameter) or previous parameter. 150/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 151
The figure below shows the first level of the configuration menu and the links to the respective second level menus (indicated below). TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 151/190...
Page 152
SERIAL: accepts commands from the serial interface (or Profibus ▪ interface if present) PROFIBUS: this operating mode is present only if the Profibus ▪ interface is installed on the controller 152/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 153
Technical Information Programming Pump Setting Menu Figure 39 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 153/190...
Page 154
Please always set up the LOAD GAS TYPE in accordance to the gas type used in the process. An improper selection of this setting could result in the damage of the pump. 154/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 155
ACTIVE STOP: enables (YES) or disables (NO) active braking. ▪ STOP SPEED READING: enables (YES) or disables (NO) the pump ▪ rotational frequency reading function also during the slowing phase that follows a stop command TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 155/190...
Page 156
START/STOP input is de- activated. The interlock input must be constantly active (INTERLOCK TYPE parameter = continuous) or may be active only at the time of starting (INTERLOCK TYPE parameter = impulsive). 156/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 157
▪ POWER: the power supplied to the pump is shown ▪ TEMPERATURE: the temperature of the pump is shown ▪ PRESSURE: the pressure read by the gauge is shown. ▪ TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 157/190...
Page 158
This setting acts only on the measurement shown on the display. Possible options: KRPM ▪ ▪ LCD CONTRAST ▪ 158/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 159
F. 1 POWER TIME NORMAL PRESSURE R1/R2 SET VALUE 0000 9999 R1/R2 HYSTERESYS R1/R2 MASK TIME 00:00:00 00:00:00 SEC 99:99:99 R1/R2 OUTPUT _|-- LOGIC: _|-- --|_ Flow 4/6 Figure 41 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 159/190...
Page 160
R1 (R2) OUTPUT LOGIC: Selects the control logic of the SETPOINT ▪ output. positive logic (the relay is energized when the measurement ▪ exceeds the reference) negative logic (the relay is energized when the measurement is ▪ below the reference) 160/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 161
▪ GAUGE MODEL ID: it shows the gauge model identified by the controller. If there are more model with the same identification the user can select the right gauge connected TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 161/190...
Page 162
The pressure indicated by the instrument is equal to: P indicated = P measured x Gas correction / 10 162/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 163
RS485 (multidrop) SERIAL ADDRESS: sets the address of the node in the case of the ▪ RS485 interface. This parameter is not visible if the RS232 is selected. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 163/190...
Select by front panel if you want to work in 232 or 485, and if necessary, the serial address. Select SERIAL from the INPUT/OUTPUT menu Open the T+ software (see orderable parts table) Ready to operate by SERIAL connections. 164/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 165
OFF = Pump under speed threshold ON = Pump over speed threshold 7, 4 R2, R1 Pump Power Monitor (10V=150W) Power value Monitor (Com) Speed value Pump Frequency Monitor (10V=81 KRPM) Figure 44 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 165/190...
This option is factory installed. The ProfiBus connector pin-out is reported in the following figure. Please in order to obtain best performance use ProfiBus certified cables. Figure 45 ProfiBus connector 166/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 167
From the software point of view, the interface can be seen as two devices (one for ProfiBus and the other for the internal link) linked together. Each device has its own state machine to manage the device operation. Figure 46 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 167/190...
Page 168
Figure 47 Tab. 17 Internal link state machine Status Meaning POWER_ON Interface set-up SRC_ADDR Search for the controller serial address IDENTIFY Identify controller model and protocol DATA_EX Regular data exchange 168/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 169
As a consequence of this, the scale and the meaning of some parameters depends directly to the specific model of Turbo Pump controller connected to the interface. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 169/190...
Page 170
Set the mantissa of the exponential value of the R1 setpoint (format X.XE-XX) when it is used for pressure monitoring (Parameter Buffer, Offset 21, = 4). The value set here is divided by 10, that means 11 = 1.1 170/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 171
= Vent & Purge Auto/Man 0 = Auto (the controller drives purge and vent valves automatically) 1 = Manual (the user can command the valves by serial line or profibus) TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 171/190...
Page 172
0 = frequency 1 = current 2 = time 3= normal operation 4= pressure = Analog Output 1 Type 0 = frequency 1 = power 2 = temperature 3= pressure 172/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 173
2 = Front panel 3 = Profibus 1100 - High Speed Value 1350 Set the high rotational speed value 1100 - Low Speed Value 1350 Set the low rotational speed value TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 173/190...
Page 174
– Short Circuit (a short circuit happened between the two of the three motor phases) – Over Voltage (too high bus voltage during braking) – Verify Line Select Not used – Power Fail (internal circuitry failure) 174/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 175
0-200 Power Motor power adsorption 0-200 Power Limit Applied Maximum allowable power 0-1500 Driving Frequency Read the rotational speed of the pump 0-1500 Target Frequency Read the target rotational speed TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 175/190...
Page 176
Bit 2 = 1: first loop of requests not ended (values in the input data are ▪ not fully coherent yet) So, if the first byte is equal to 0, the interface is working properly, otherwise there is a problem. 176/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 177
ProfiBus external diagnostic service is required and some bits in the diagnosis frame signals which parameter is wrong (i.e. over range or under range). TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 177/190...
Page 178
(slave is parameterised by another master) Reserved Master Add (master address after parameterisation. FF= without parameter) Ident number high Ident number low External diagnosis “Header” length indication including header 178/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 179
Not used – Start/Stop diag status – Low Speed diag status – Soft Start diag status – Vent Valve – Purge Valve – Mode – High speed – Low Speed TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 179/190...
If within each step of the soft start mode the rotational speed of the pump do NOTE not reach the planned value within 15 minutes the display will be as shown in the following figure. 180/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 181
The message will stay on until the temperature decreases below threshold value. Press the STOP RESET pushbutton twice to start the pump. TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 181/190...
Page 182
Press the START push-button twice to start the pump. Should the message still be present, call the Agilent service. Check connections and shortages between pump and controller, then press the STOP RESET pushbutton twice to start the pump. 182/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Vent Valve N.O. 0,5 mm orifice (0,5 m cable) 9699844 Vent Valve extention cable (5 m) 9699941 Mains cable NEMA plug (3 m) 9699958 Mains cable european plug (3 m) 9699957 RS232 Serial to B/T Adapter X3514-68003 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01 183/190...
Page 184
Technical Information Orderable Parts Page intentionally left blank 184/190 TwisTorr 305 FS Remote Controller User Manual / 87-901-059-01...
Page 185
Vacuum Products Division Dear Customer, Thank you for purchasing an Agilent vacuum product. At Agilent Vacuum Products Division we make every effort to ensure that you will be satisfied with the product and/or service you have purchased. As part of our Continuous Improvement effort, we ask that you report to us any problem you may have had with the purchase or operation of our products.
Page 187
Please follow these instructions whenever one of our products needs to be returned. Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to include the completed Health and Safety declaration Section. No work can be started on your unit until we receive a completed copy of this form.
Page 188
Terms and conditions TERMS AND CONDITIONS Please read the terms and conditions below as they apply to all returns and are in addition to the Agilent Technologies Vacuum Product Division – Products and Services Terms of Sale. Unless otherwise pre-negotiated, customer is responsible for the freight charges for the returning product.
Page 189
Vacuum Products Division Request for Return Form Customer information Company : Contact Name: Address: Tel: Fax: Email: Equipment Product description Agilent PartNo Agilent Serial No Original Purchasing Reference Failure description Type of process (for which the equipment was used) Type of return Non Billable Billable New PO # (hard copy must be submitted with this form): ______________________________________...
Page 190