If the display does not accept the 500 MHz cw
frequency data input
Check that the HP 70341A HP-MSIB row address is greater than the HP
70340A row address.
Check that the HP 70341A HP-MSIB column address is equal to (or greater
than) the HP 70340A column address.
Use the
view each module's HP-MSIB address using the front panel screen.
If the HP-MSIB address switches are not set correctly:
1. Correctly reset the HP-MSIB addresses so that the HP 70340A can address
the HP 70341A.
Refer to the procedure, \To Check and Set the HP 70341A HP-MSIB
Address" for help in determining or setting the HP-MSIB addresses.
2. Cycle power.