13
CLOSING DOWN
Note
No special measures are to be taken in the event of short-term shutdowns of the gas conditioning
system.
In the case of long-term shutdowns, for example after a completed series of measurements, it is
recommended to purge the gas conditioning system with fresh air or inert gas. A flushing time of 3 to
5 minutes is sufficient under normal conditions. Condensate residues must also be removed from the
system.
Warning
24
The installation site of the gas conditioning unit must remain frost-
free even when the unit is switched off.
Aggressive condensate is possible.
Wear protective glasses and proper protective clothing!
PSS-20 | 1.00.01