① Peltier temperature controller TC1
② Filter FPF+ with 2 μm filter porosity(including: LA3 humidity sensor)
③ Peristaltic pump SR25.2 for condensate removal
④ Acid dosing Inlet (optional)
⑤ Gas sample line temperature controller TC2 (optional)
⑥ Flow meters 2 x FM40 with sample gas outlet (measuring range: 15 – 150 l/h & 25 – 250 l/h)
⑦ Filter FPF with 0.1μm filter porosity
⑧ Acid dosing peristaltic pump SR25.3 (optional)
⑨ Peristaltic pump SR25.2 for condensate removal
⑩ Condensate outlet
Figure 3
Design of the conditioning unit PSS-20, front-view
In order to prevent the liquid from entering the downstream analytical instrument and improve the
working safety of the entire system, the humidity sensor LA3 is installed in the FPF+ filter as
standard component.
When the humidity sensor detects that the liquid enters the filter, the system will automatically turn
off the sample gas pump. When the liquid is cleaned, the system will return to normal operation.
Filter FPF+ Fig. 3/②
adequate particle separation.
www.mc-techgroup.com
and FPF Fig. 3/
⑦ are installed in front of diaphragm pump
PSS-20 | 1.00.01
Fig. 2/③
to ensure