Indication
Problem
The system
Peltier
works
cooling unit
normally
and pump
are
operating
normally.
There is
condensate
in the gas
tubing.
www.mc-techgroup.com
Possible Cause
Peristaltic pump
failure
LA humidity sensor
failure
PSS-20 | 1.00.01
Check/Solution
Replace the pump tubing.
Check the peristaltic pump.
(See manual SR25.2 manual for manual
peristaltic pump)
Check the LA humidity sensor
27