15

TROUBLE SHOOTING

The following table aims to point out possible operational problems and offer solutions to such
problems (not applicable during the starting procedure).
Indication
Problem
The power
Device can
light doesn't
not start
work
Temperature
Display
controller
9999 or
display error
1999
Sample
Peltier
pump does
cooling unit
not start
worked
normally,
but there
was no gas
flow
Small outlet
Peltier
flow
cooling unit
and pump
are
operating
normally,
but the
outflow rate
is very low
26
Possible Cause
No voltage
Incorrect voltage
Fuse broken
Bad contact of
PT100 sensor
signal wire;
PT100 sensor
signal wiring error;
PT100 sensor is
damaged;
Diaphragm pump
failure;
Diaphragm pump
contamination;
The gas sample
probe/gas line is
blocked or
compressed;
The sample gas
line connected to
the analyzer is
blocked or
compressed;
The sample pump
switch is not
turned on;
Humidity sensor
contact switch to
close the sample
pump;
Filter clogging
Pump failure
PSS-20 | 1.00.01
Check/Solution
Check the power supply voltage against
the model's name plate;
Check if the power plug is inserted
correctly and that the main switch is in
position "1".
Check the fuse in the power socket.
Replace the spare fuse;
Open the top cover to check the
thermostat PT100 sensor wiring.
Check if the wiring is correct.
Replace the PT100 sensor;
Remove the gas lines from the pump head
and check it;
Clean the pump as needed.
Remove the gas tubing from the gas inlet;
Clean or change the gas way;
Take down the gas tubing on the analyzer
side. If the sample gas is flowing, check
the connection of the gas tubing.
Clean or change the gas way;
Open the sample gas pump switch.
Check the peristaltic pump discharge.
Check the condensate discharged from the
tube.
Check the filter, if the inside is wet,
replace the filter element in time, and dry
the filter chamber.
Check the cooler according to the
instruction manual.
Adjust the desired flow rate with a needle
valve.
Clean the filter and replace the filter
element.
Check the pump.
www.mc-techgroup.com