2.3 Installing the Flowcell – Figs. 2.4 and 2.5
Notes.
Connecting pipework may be flexible plastic or rigid PVC, polypropylene or metal depending on the installation.
Isolating valves should be fitted to allow removal of the instrument.
Space should be left on each side of the assembly to allow access to the sensors.
For maintenance purposes the following minimum clearances are recommended:
Dimensions in mm
118
Receiver
Sample inlet
(12 mm i.d. flexible
hose connection).
Fig. 2.4 Overall Dimensions and Mounting Details of 7320 100 (Low Range Sensor)
Left
(for receiver removal)
Right (for emitter removal
Top
(for cleaner removal)
327
155 CRS
Cleaner
Sample outlet,
see Fig. 2.7
for location.
Four holes ∅6
for mounting
Drain
(12 mm i.d. flexible
hose connection)
2 MECHANICAL INSTALLATION...
100 mm
100 mm
200 mm
Emitter
410
.
Sample Outlet
Connector
for 6 mm i.d.
Tube Located
Behind the
Sensor Body
5