5.6 Set Up Outputs

Set Up
Parameter
Set Up Outputs
Current o/p Span
10.00 mg/l C
Default Output
8.00 mg/l C
Current o/p Type
4 to 20 mA
Test Current o/p
0.00 mg/l C
A1 Action
Off
A1 Setpoint
1.50 mg/l C
A2 Action
Off
A2 Setpoint
2.50 mg/l C

5.7 Factory Settings

Set Up
Outputs
Factory Settings
Factory Set Code
0000
Set Up Outputs Page
Factory
Settings
Set the current output span between the following limits:
Low range sensor: 0 to 2 and 0 to 20 mgl
High range sensor: 0 to 20 and 0 to 100 mgl
Set the current output to a default in the event of a diagnostic alarm.
Select the current output:
4 to 20mA
0 to 20mA
0 - 20
0 to 10mA
0 - 10
Set the value within the current o/p span.
The instrument automatically transmits a test signal to represent the current
output range.
Alarm 1 Action
This can be set as a high or low alarm contact.
Select High or Low as appropriate.
Low
Select Off to disable this alarm.
High
Adjust setpoint to a value within the range of the sensor.
Alarm 2 Action
This can be set as a high or low alarm contact.
Select High or Low as appropriate.
Low
Select Off to disable this alarm.
High
Adjust setpoint to a value within the range of the sensor.
Return to top of
These parameters are set at the factory and will not normally require
adjustment on site. However, if the receiver/emitter modules have been
changed then the emitter light will need setting. The full Factory Settings
Diss. Organics
page
page is given in Section 6.4.
.
Set Up Outputs Page
5 PROGRAMMING
–1
C
–1