Operating Display Page - ABB 7320 Series Instruction Manual

Dissolved organics monitor
Table of Contents

Advertisement

...5 PROGRAMMING

5.2 Operating Display Page

Note. This is the default page. The programme returns to this point from any of the programming pages if no data has
been entered after four minutes.
Diss. Organics
mg/l
D. Org
8.5 mg/l C
Alarm 1 Setpoint
1.5 mg/l
Alarm 2 Setpoint
8.6 mg/l
Lamp Disabled
Manual Clean
14
or
°H
Sensor
Calibration
No
Yes
No
Yes
Dissolved Organics Mode Operating Page
Display in
(see Appendix)
inferred units
Absorbance/metre; colour °H; Coag. Dose mgl
Equivalent value in mgl
–1
Dissolved Organics
This is for display only. See
This is for display only. See
Switching the Lamp On/Off
In the interests of safety it is essential that the lamp is switched off before
doing any maintenance on the sensor. When off, Lamp Disabled is
displayed in the lower window of the Diss. Organics page; the top
window will be blank.
Press the
switch to disable the lamp and
Manual Cleaning
switch to change No to Yes and press the
Press the
a manual clean.
Return to top of
Operating Display Page
–1
–1
; mgl
.
for programming details.
Set Up Outputs Page
for programming details.
Set Up Outputs Page
to switch the lamp back on.
.
T.O.C.
switch to start