ABB 7320 Series Instruction Manual page 8

Dissolved organics monitor
Table of Contents

Advertisement

...2 MECHANICAL INSTALLATION
Dimensions in mm
162
110
Bracket Fixed to
Flowcell Cover,
Emitter side
Fig. 2.5 Overall Dimensions and Mounting Details of 7320 200 and 7320 300 (High Range Sensors)
6
For maintenance purposes the following minimum
clearances are recommended:
Left
(for receiver removal)
Right (for emitter removal
Top
(for filling with standard solution)
Cleaner
373
Receiver
Mounting Bracket
(4 holes O6)
two parts to facilitate emitter module removal during
maintenance. See Section 6 - Maintenance .
Fix to Wall
150 mm
100 mm
200 mm
405
155 CRS
Note. The emitter end mounting bracket is in
Standard Solution
Filler
Sample Outlet
(for 8 mm i.d. Tube)
Emitter
Sample Inlet
(12 mm i.d. flexible
.
hose connection)
Drain
(12 mm i.d. flexible
.
hose connection)