...5 PROGRAMMING

5.5 Set Up Parameters

Security
Code
Set Up Parameter
Sensor Type
7320 100
Damping Value
5
Inferred Units
Diss. Organics
Absorbance/metre
Colour °H
Coag. Dose mg/l
mg/l T.O.C.
Conversion Fac.
2.00
Clean Int.
60 min
Flow Alarm
Off
Normally Open
Normally Closed
Alter Sec. Code
0000
Alter Cal. Code
0000
16
Set Up Parameters Page
Set Up
Outputs
Select the appropriate sensor:
7320 200
7320 300
When changed Please Wait is displayed for 20 seconds.
Enter a damping value in the range 1 to 20.
This is used to prevent short term variations in reading, typically due to
bubbles in the sample. Always use the
acceptably stable reading.
Dissolved Organics mode
Inferred Units mode
See Appendix.
Enter the conversion factor between Dissolved Organics and the Inferred
Unit selection.
Cleaning Interval
Enter required interval between automatic cleaning procedures.
Options: 15, 30, 45 & 60 minutes. 2, 4, 6, 12 & 24 hours.
Flow Alarm Input Configuration
Set the normal 'none' alarm condition (normally open or closed), or disable
by setting to Off.
Alter Set Up Parameter Security Code
Enter value in the range 0000 to 9999.
Alter Calibration Security Code
Enter value in the range 0000 to 9999.
Return to top of
7320 100 = Low Range Sensor.
7320 200 = High Range Sensor.
7320 300 = Coagulation Dose Sensor.
Note. Absorbance/metre has a fixed factor of 1.5.
Set Up Parameters Page
value which gives an
lowest
.