ABB 7320 Series Instruction Manual page 22

Dissolved organics monitor
Table of Contents

Advertisement

...6 MAINTENANCE
Unscrew the collar and carefully
withdraw the module.
1
Optical
Receiver
Split the flowcell to access the flow chamber
3
– see Fig. 6.3 for procedure.
Clean the inside of the flow chamber
4
and other assemblies thoroughly.
Use mild detergent and rinse with de-
ionized water.
Inspect the wiper blades for wear or
damage and fit new ones if necessary
– see Fig. 6.4 for assembly details.
Reassemble flowcell using new seals.
5
Ensure that the modules locate in the keways before tightening the collars.
20
Sample Outlet
Unscrew the collar and carefully
2
withdraw the module.
Inlet/Isolating
Valve
Drain
Valve
Fig. 6.2 High Range Sensors – Cleaning the Flowcell
Optical
Emitter
Seal
Seal