ABB 7320 Series Instruction Manual page 23

Dissolved organics monitor
Table of Contents

Advertisement

Remove the four stainless steel
closing screws holding the flowcell
halves together.
Thread two of the stainless steel screws into
holes provided and advance them slowly and
evenly to push the cell halves apart. When
the O-ring seal is clear of the body the cell
halves should separate easily.
Vew from emitter end
2
3
Fig. 6.3 Splitting the High Range Flowcell
6 MAINTENANCE...
Remove the securing screws to free
1
the mounting plate from the angle
bracket (2).
Thoroughly check and clean the inside
4
of the cell. The wiper blades can also
be serviced/changed – see Fig. 6.4.
Reassemble.
5
21