• After a manual abortion, the C 150 reverts to its standard operational status. The indication " - - - " in
the lower display shows that the CO
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing "EXIT".
• If necessary, repeat the sterilization.
• When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
sensor.
2
The CO
• Turn on the unit (chap. 5).

15.5.2 Aborting the hot-air sterilization by opening the outer door

For safety reasons, the sterilization cycle is automatically aborted if you open the outer door of the C 150.
• Do not open the glass door and close the outer door immediately.
• After an abortion caused by opening the outer door, the C 150 reverts to its standard operational
status. The indication " - - - " in the lower display shows that the CO
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing "EXIT".
• If necessary, repeat the sterilization.
• When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
sensor.
2
The CO
• Turn on the unit (chap. 5).

15.5.3 Aborting the hot-air sterilization using the main power switch

For safety reasons, the sterilization cycle is aborted if the C 150 is turned off or if there is a power failure.
Normally you should not terminate sterilization in this way.
• Do not open the doors of the C 150 until the interior temperature has dropped to 37 °C / 98.6°F.
• After turning on again the C 150, it reverts to its standard operational status. The indication " - - - " in
the lower display shows that the CO
As long as the interior temperature remains above the temperature set to trigger the temperature
alarm, the C 150 will trigger alarm signals. You can ignore this alarm and turn off the buzzer of the
safety controller by pressing "EXIT".
• If necessary, repeat the sterilization.
• When the inner chamber has cooled down to a value below 60 °C / 140°F, turn off the unit and plug in
the CO
sensor.
2
The CO
• Turn on the unit (chap. 5).
C 150 (E2) 06/2011
sensor is disconnected.
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
2
sensor is disconnected.
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140°F.
2
sensor is disconnected.
2
page 64/78