The CO
incubators were produced in accordance with VDE regulations and were routinely tested in ac-
2
cordance to VDE 0411-1 (IEC 61010-1).
The glass door and the inner chamber will become hot during operation.
Danger of burning.
∅ Do NOT touch the glass door, the inner surfaces or the charging material during opera-
tion.
Stability hazard.
Danger of injury.
Damage to the unit and the charging material.
Housing cover breakaway.
∅ Do NOT climb on the lower housing cover.
∅ Do NOT load the lower housing cover with heavy objects while the unit door is open.
Vent out any gas that may escape via good room ventilation or a suitable exhaust system.
High concentration of CO
Danger of poisoning.
∅ Do NOT set up units in non-ventilated recesses
Ensure technical ventilation measures
Observe the relevant regulations for handling CO
1.6
Intended use
Series C 150 incubators are suitable for the cultivation of mammal cells under typical conditions of
approx. 37 °C / 98.6 °F. The incubator permits setting defined pH conditions by common NaHCO
systems of commercial cell media by keeping an exact CO
tee high humidity inside to avoid osmolarity increasing caused by the evaporation of the cell media.
Observing the instructions in this operating manual and conducting regular maintenance work
(chap. 16) is part of the intended use.
Other applications are not approved.
WARNING: If customer should use a BINDER chamber running in non-supervised continu-
ous operation, we strongly recommend in case of inclusion of irrecoverable specimen or
samples to split such specimen or samples and store them in at least two chambers, if this is
feasible.
Always use proper protective equipment (clothing, gloves, safety glasses, etc.).
Always follow good hygienic practices according to GLP/SOP protocols.
Each individual operating the C 150 incubator is responsible for his or her own safety
C 150 (E2) 06/2011
CAUTION
WARNING
WARNING
.
2
atmosphere inside. C 150 incubators guaran-
2
.
2
buffer
3
page 13/78
Need help?
Do you have a question about the APT.line C 150 and is the answer not in the manual?