2.5

Unit doors

The outer door of the C 150 is equipped with a heater on its inner side. The door must be closed while
the unit is operating normally in order to ensure stable climatic conditions in the inner chamber.
An additional glass door enables viewing of the samples without disturbing the temperature in the interior
and contaminating the samples sealing the interior of the C 150.
When the outer door is open, the CO
Delay time for the temperature and CO
After closing the outer door, the tolerance range alarm is turned off for a programmable delay
time. This prevents alarms being constantly triggered during the unstable operating phase
after opening the outer door.
3.

Completeness of delivery, transportation, storage, and installation

3.1

Unpacking, and checking equipment and completeness of delivery

After unpacking, please check the unit and its optional accessories, if any, based on the delivery receipt
for completeness and for transportation damage. Inform the carrier immediately if transportation damage
has occurred.
The final tests of the manufacturer may have caused traces of the shelves on the inner surfaces. This
has no impact on the function and performance of the unit.
Please remove any transportation protection devices and adhesives in/on the unit and on the doors and
remove the operating manuals and accessory equipment.
For transport purpose, a silica gel bag for drying purpose was added. Do not eat! Do not open the silica
gel bag and dispose of it with normal waste.
Remove any protective lamination sheet on the inner metal surfaces prior to commissioning.
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
∅ Do NOT lift or transport the unit using the door handle, the door or the lower housing.
Lift the unit from the pallet at the four lower corners with the aid of four people.
If you need to return the unit, please use the original packing and observe the guidelines for safe lifting
and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 17.1.
Note on second-hand units (Ex-Demo-Units):
Second-hand units are units that were used for a short time for tests or exhibitions. They are thoroughly
tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly.
Second-hand units are marked with a sticker on the unit door. Please remove the sticker before commis-
sioning the unit.
C 150 (E2) 06/2011
intake valve automatically closes.
2
tolerance range alarm :
2
CAUTION
page 17/78