Display Setting range
OFF, or -199,9 °C
to +300 °C
OFF, or
-10.0 vol.-% to
110.0 vol.-%
The operating function "At.S" specifies the safety controller set point type "Limit" (Lit) or "Off-
set" (OFS).
• With set point type "Limit" (Lit), operating function "AL.S" (safety controller set point as limit
value) is displayed in operating modes HAND and USER. You can set a limit value .
• With set point type "Offset" (OFS), operating function "Ot.S" (safety controller set point as
offset value) is displayed in operating modes HAND and USER. You can set an offset
value.
C 150 (E2) 06/2011
Operating function
Firmware revision of safety controller for service / maintenance pur-
poses.
Data record (year) for service / maintenance purposes.
Data record (month) for service / maintenance purposes.
Data record (day) for service / maintenance purposes.
Data set version for service / maintenance purposes.
Temperature adjustment serves to adjust the temperature sensor.
final value adjustment serves to adjust the CO
CO
2
sensor.
2
page 34/78