11 Supplement
Ʋ Channel 9
Transmission rate
Current value
Resolution, digital
Output variable - X-ray alarm
Output
Ʋ Signal
Ʋ Physical layer
Damping (63 % of the input variable)
Channel Numbers
Ʋ Channel 1
Ʋ Channel 8
Ʋ Channel 9
Transmission rate
Current value
Resolution, digital
Output variable - Real value correction
Output
Ʋ Signal
Ʋ Physical layer
Damping (63 % of the input variable)
Channel Numbers
Ʋ Channel 1
Ʋ Channel 8
Ʋ Channel 9
Transmission rate
Current value
Resolution, digital
Relay output
Output
Switching voltage
Switching current
Switching current
Ʋ Standard
Ʋ USA, Canada
Breaking capacity
Ʋ Min.
76
Pulse rate
31.25 Kbit/s
10 mA, ±0.5 mA
> 0.1 mm (0.004 in)
digital output signal, Foundation Fieldbus protocol
according to IEC 61158-2
0 ... 999 s, adjustable
Process value
Electronics temperature
Pulse rate
31.25 Kbit/s
10 mA, ±0.5 mA
> 0.1 mm (0.004 in)
digital output signal, Foundation Fieldbus protocol
according to IEC 61158-2
0 ... 999 s, adjustable
Process value
Electronics temperature
Pulse rate
31.25 Kbit/s
10 mA, ±0.5 mA
> 0.1 mm (0.004 in)
Relay output (SPDT), floating spdt
max. 253 V AC/DC
With circuits > 150 V AC/DC, the relay contacts must be
in the same circuit.
max. 3 A AC (cos phi > 0.9), 1 A DC
max. 3 A AC (cos phi > 0.9), 1 A DC
max. 3 A AC (cos phi > 0.9)
50 mW
MINITRAC 31 • Foundation Fieldbus