6 Set up with the display and adjustment module
Date/Time
Reset
56
6.5.4 Additional adjustments
In this menu item you can set the actual date, time and display format.
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
The following reset functions are available:
Basic settings: Resetting of the parameter adjustments to default
values at the time of shipment. Order-specific settings are deleted.
Default settings: Resetting of the parameter adjustment like under
" Basic settings". In addition, special parameters are reset to default
values. Order-specific settings are deleted.
Peak values of measured value: Resetting of the parameter adjust-
ments in the menu item " Setup" to the default values of the respec-
tive instrument. Order-specific settings remain but are not taken over
into the current parameters.
Peak values of temperature: Resetting of the measured min. and
max. temperatures to the actual measured value.
The following table shows the default values of the instrument. The
values apply for the application " Limit level". First of all you have to
select the application.
Depending on the instrument version, not all menu items may be
available or they may be differently assigned:
MINITRAC 31 • Foundation Fieldbus