Adjustment data
MINITRAC 31 • Foundation Fieldbus
parameters to ensure the optimum adjustment of the display. The
procedure is described in the following.
This menu item enables the setting of the requested national lan-
guage.
In delivery status, the sensor is set to the ordered national language.
If no language is preset, you will be asked during setup.
With this parameter you can change the indication of the display.
You can choose if the display should show the actual pulse rate, the
electronics temperature or the percentage value.
You can choose whether the display should show the current pulse
rate, the electronics temperature, the process value or the output
signal.
6.4.3 Diagnostics
In this menu item, you can enquire the status of your sensor. In normal
operation, the sensor displays the message " OK". In case of fault,
you will find the corresponding fault code here.
The peak value function holds the max. and min. values during opera-
tion.
Pulse rates - min./max.
Temperature - min./max./actually
Here you can retrieve the adjustment value of the sensor. This is the
percentage value of the difference of the min. and max. adjustment
points (Delta I). The value is an indication for the reliability and non-
repeadability of the measurement.
The higher the difference between the two adjustment points,
the higher the differential value (Delta I) and the more reliable the
measurement. A Delta I value below 10 % is an indication for a critical
measurement.