Vega MINITRAC 31 Operating Instructions Manual page 61

Radiometric sensor for density measurement
Table of Contents

Advertisement

Main menu
Procedure
Isotope
MINITRAC 31 • Foundation Fieldbus
You have to enter the addresses of the Secondary instruments in the
list of the Primary instruments. This function is not possible in the sup-
ply and adjustment module. To do this you need PACTware with the
respective DTM.
The main menu is divided into five sections with the following func-
tions:
Setup: Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Additional adjustments: Instrument unit, reset, date/time, copying
function
Info: Instrument name, hardware and software version, date of manu-
facture, instrument features
Check if the correct language is already set for the display. If not,
you can change the language in the menu item " Display - Menu
language".
Start with the setup of MINITRAC 31.
In the main menu item " Setup", the individual submenu items should
be selected one after the other and provided with the correct param-
eters to ensure optimum setting of the measurement. The procedure
is described in the following.
Stick with the normal sequence of the menu items as closely as pos-
sible.
6.7.1 Setup
In this menu item you can adjust the MINITRAC 31 to the isotope
installed in the source container.
For this purpose, check which isotope is in the source container. You
can find this information on the type label of the source container.
With this selection, the sensitivity of the sensor is adapted in an opti-
mum way to the isotope.
6 Set up with the display and adjustment module
61