Vega MINITRAC 31 Operating Instructions Manual page 82

Radiometric sensor for density measurement
Table of Contents

Advertisement

11 Supplement
FF descriptor
ANALOG_IN
FF_CHANNEL_AVAIL-
ABLE
FF_CHANNEL_USED
Mapping of Process Value Status
Hex
Quality
0x00
Bad
0x01
Bad
0x0C
Bad
0x10
Bad
0x1C
Bad
0x20
Bad
0x44
Uncertain
0x48
Uncertain
0x4C
Uncertain
0x80
Good (NC)
Mapping of User Error Codes to FF Field Diagnostics
Priority
Description
31
Hardware failure
30
Memory failure
29
Software failure
28
Paramererization corrupt
27
Undefined 27
26
Undefined 26
82
Rel. In-
Description
dex
58
Current on analog input
59
Available channels
60
Used channels
Sub-Status
Non-specific
Configuration Error
Device Failure
Sensor Failure
Out of Service
Transducer in MAN
Last Usable Value
Substitute
Initial Value
Non-specific
Condition
Unexpected error
It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,
117, 120
Five or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,
86, 121, 122, 141
On user error codes: 40, 41, 53, 68, 123, 124, 125
Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application
Transducer block is in mode "Manual"
Three or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 29
After startup of device or channel assignment in
AIFB was changed. The channel unit may be un-
known until next "Process Data Update"
On user error codes: 13
No errors concerning to channel handling
On user error codes: 0, 33, 45, 71, 126, 127
User Error Codes
40, 41
34, 35, 36, 37
80
72
-
-
Unit
-
-
-
NE-107 Status
FAILURE
FAILURE
FAILURE
FAILURE
-
-
MINITRAC 31 • Foundation Fieldbus