Operating menu

Device tag

Navigation
Description
Factory setting

Distance unit

Navigation
Description
Selection
Factory setting

Tank type

Navigation
Prerequisite
Description
Selection
Factory setting
116
16.3

"Setup" menu

: Marks the navigation path to the parameter via the display and operating
module.
: Marks the navigation path to the parameter via an operating tool (e.g.
FieldCare).
: Marks parameters which can be locked via the software locking →  55.
Navigation
 Setup
 Setup → Device tag
Enter a unique name for the measuring point to identify the device quickly within the
plant.
FMP5x
 Setup → Distance unit
Used for the basic calibration (Empty / Full).
SI units
• mm
• m
m
 Setup → Tank type
Medium type (→  127) = Liquid
Select tank type.
• Metallic
• Bypass / pipe
• Non metallic
• Mounted outside
• Coaxial
Depending on the probe
US units
• ft
• in
Levelflex FMP53 HART