16.4
"Diagnostics" menu
Navigation
Diagnostics
Diagnostics → Actual diagnostics
Displays current diagnostic message.
The display consists of:
• Symbol for event behavior
• Code for diagnostic behavior
• Operating time of occurrence
• Event text
If several messages are active at the same time, the messages with the highest
priority is displayed.
Information on what is causing the message, and remedy measures, can be viewed via
the
symbol on the display.
Diagnostics → Timestamp
Displays the timestamp for the currently active diagnostic message.
Diagnostics → Previous diagnostics
Displays the last diagnostic message which has been active before the current message.
The display consists of:
• Symbol for event behavior
• Code for diagnostic behavior
• Operating time of occurrence
• Event text
The condition displayed may still apply. Information on what is causing the message,
and remedy measures, can be viewed via the
Levelflex FMP53 HART
symbol on the display.