Test Mode Operating Mode
Example 9. BCH Test Function Using Manual Receiver Control
The following example shows how to set up a test mode measurement using the BCH test function. In this
example the test set is configured to transmit a BCH on PGSM channel 21. Manual receiver control is used to
configure the test set to measure a signal from the mobile station at 895 MHz and 14 dBm. Note that the
frequency can also be tuned by channel number.
1. Select test mode.
OUTPUT 714;"CALL:OPERATING:MODE TEST"
2. Select PGSM as the broadcast band.
OUTPUT 714;"CALL:CELL:BAND PGSM"
3. Select BCH as the test function.
OUTPUT 714;"CALL:FUNCTION:DOWNLINK BCH"
4. Set the BCH to channel 21.
OUTPUT 714;"CALL:BCH:PGSM 21"
5. Configure the receiver control to manual.
OUTPUT 714;"RFANALYZER:CONTROL:AUTO OFF"
6. Configure the test set's receiver frequency to 895 MHz.
OUTPUT 714;"RFANALYZER:MANUAL:FREQUENCY 895 MHZ"
If tuning by channel number, see
7. Set the receiver's expected power level to 14 dBm.
OUTPUT 714;"RFANALYZER:EXPECTED:POWER 14 DBM"
8. To make the measurement, set up the mobile station to transmit at 895 MHz and at a power level of 14
dBm.
528
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\gen_bse_test.fm
"RFANalyzer:MANual:CHANnel[:SELected]" on page 375