READ:ORFSpectrum[:ALL]?
Function
Queries (initiates and fetches) one output RF spectrum measurement as a sequential operation.
Returns Integrity Indicator see
Spectrum due to Switching (Max), 30 kHz Bandwidth Power, and Output RF Spectrum due to
Modulation (Average). The FETCh command should be used to obtain other measurement
results. See
Query
Integrity indicator
• Range: 0 to 16
• Resolution: 1
TX power
• Range: −100 to +100 dBm and 9.91E+37 (NAN)
• Resolution: 0.01 dB
Output RFspectrum due to switching (Max)
• Range: 0 to 8 comma separated values −100 to +100 dBm and 9.91E+37 (NAN)
• Resolution: 0.01 dB
30 kHz bandwidth power
• Range: −100 to +100 dBm and 9.91E+37 (NAN)
• Resolution: 0.01 dB
Output RF spectrum due to modulation (average)
• Range: 0 to 22 comma separated values −200 to +100 dBm and 9.91E+37 (NAN)
• Resolution: 0.01 dB
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
"Integrity Indicator" on page
"FETCh:ORFSpectrum" on page
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_read.fm
125, TX Power, Output RF
322.